Table 3.

Limit of accuracy for AQ analysis

RNA copy number/mL Proportion of PA/I38T (%) PyroMark ID quality score
107 36.6 ± 0.8 Pass
106 36.6 ± 0.9 Pass
105 38.0 ± 2.5 Pass
104 38.6 ± 0.4 Pass
103 39.3 ± 0.2 Pass
102 59.2 ± 2.0 Check
101 27.2 ± 25.7 Fail

PA/I38T proportions are the average ± standard deviation of three experiments.