TABLE 2.
Field of view (mm3) | 100*100*60 | 100*100*48 | 125*125*170 | 200*200*50 | |
Resolution (mm) | 1.56*1.56*3 | 1.5*1.5*4 | 2*2*5 | 2.5*2.5*5 | |
Matrix size | 64*64 | 68*68 | 64*64 | 80*80 | |
Flip angle (°) | 12 | 8 | 8 | 8 | |
Triggering frequency (mm−1) | 72 | 34 | 40 | Varies | |
Number of phases | 16 | 15 | 18 | 16 | |
Cartesian readout | TE/TR (msec) | 4.3/12 | 4.2/14 | 3.6/11 | 2.8/Varies |
Readout time (msec) | 3.4 | .3.5 | 3.1 | 2.1 | |
Systole Venc (cm/s) | 400 or 250 | 300 | 200 | 400 | |
Diastole Venc (cm/s) | 100 | — | 140 | — | |
Venc switch time (msec) | 340 | — | 800 | — | |
Scan time(mm:ss) | 27:46 | 9:44 | 70:25 | 7:27 | |
Spiral readout | TE/TR (msec) | 2.3/12 | 2.4/14 | 2/11 | 2.3/Varies |
Readout time (msec) | 4 | 4 | 4 | 4 | |
Numbet of interleaves | 36 | 32 | 32 | 16 | |
Systole Venc (cm/s) | 400 or 250 | 300 | 200 | 400 | |
Diastole Venc (cm/s) | 100 | 90 | 140 | 100 | |
Venc switch time (msec) | 340 | 850 | 800 | Varies (default = 400) | |
Scan time (mm:ss) | 15:38 | 7:05 | 35:14 | 3:00 |
In each case, the Venc was specified via a prior 2D through plane phase-contrast scan. In the table Qmax represents the maximum rate in the prescribed flow waveform. Please note that the scan time for human subjects stated in this table do not take into account respiratory gating, which in general increase the stated times by 3 times or more (depending on individual subjects’ gating efficiency).