Skip to main content
. 2020 Jun 17;15:225. doi: 10.1186/s13018-020-01736-0

Table 1.

Precision of radiostereometric analysis for assessment of stem migration

Axis Translation (mm)* Rotation (°)*
Transverse (x) 0.23 0.46
Longitudinal (y) 0.18 1.14
Sagittal (z) 0.30 0.24

*Precision of measurements based on 75 double investigations. Given number represents the smallest migration value that is considered significant and is based on 2 standard deviations of the error obtained. This, hence, represents the 95% confidence limit