Skip to main content
. Author manuscript; available in PMC: 2020 Jun 18.
Published in final edited form as: Med Phys. 2019 Sep 23;46(11):4881–4897. doi: 10.1002/mp.13805

TABLE II.

Comparison of experimental and theoretical σcbv* of the three canine subjects used in the Part I paper.1 The values after the ± sign are standard deviation of data measured at 8 slice locations. The units of σcbv* are (ml/100 g).

Experimental Theoretical
Canine Subject Number σcbv* (constant mA) σcbv* (modulated mA) % noise reduction σcbv* (constant mA) σcbv* (modulated mA) % noise reduction
1 2.17 ± 0.12 1.32 ± 0.12 39% ± 7% 2.18 1.35 38%
2 2.24 ± 0.35 1.60 ± 0.18 29% ± 13% 2.24 1.64 27%
3 1.99 ± 0.14 1.33 ± 0.06 33% ± 6% 1.94 1.32 32%