. 2020 Jun 18;10:9906. doi: 10.1038/s41598-020-66899-6

Table 1.

SapienXT patients selected for the study.

Patient Age at TAVI Gender Volume of leaflet calcium [mm3] AV peak gradient [mmHg] SapienXT size [mm] Fluoro Diam [mm] FE-Fluoro Diameter Difference [mm] (%) Paravalvular leakage
S-1 85 M 385 46 29 25.3 1.0 (3.8) trivial in NCC
S-2 77 M 627 48 29 27.1 −0.7 (−2.7) trivial in NCC
S-3 78 M 220 67 26 26.3 −0.4 (−1.6) trivial in NCC
S-4 59 F 1121 67 26 23.2 0.8 (3.6) trivial in RCC
S-5 69 F 690 70 26 23.7 1.8 (7.3) 1 trivial jet in RCC-LCC, 1 trivial jet in NCC
S-6 76 M 751 70 26 25.0 −0.1 (−0.2) No PVL
S-7 85 M 527 72 26 25.1 0.6 (2.5) trivial in RCC, almost absent
S-8 78 M 1997 72 26 25.4 0.9 (3.3) 1 mild jet in RCC, 1 trivial jet in NCC
S-9 88 F 380 73 23 21.4 0.8 (3.5) 1 trivial jet LCC-NCC, 1 mild+ jet RCC-NCC
S-10 83 M 1478 76 29 25.5 0.5 (1.8) trivial in NCC-RCC
S-11 78 F 221 78 26 22.7 2.7 (11.2) trivial in RCC-LCC
S-12 83 M 553 83 26 25.6 −0.6 (−2.3) Trivial
S-13 81 M 824 100 29 26.4 −0.1 (−0.3) 1 mild jet in RCC, 1 trivial jet in RCC-NCC
S-14 90 F 385 103 26 23.5 1.3 (5.5) No PVL
Average + /− std 79.3 ± 8.0 726 ± 503 73 ± 16 24.7 +/− 1.6 0.6 + /− 0.9 (2.5 + /− 3.9)

The last three columns report the stent diameter measured from fluoroscopy, the % difference of this diameter from computational results and the clinical outcome in terms of paravalvular leakage (PVL). RCC = right coronary cusp; LCC = left coronary cusp; NCC = non-coronary cusp.