Table 4.

Max Principal Strains on the implantation site for the cohort of 14 patients who underwent the implantation of CoreValve device.

Max Principal Strains in the implantation site [%]
Patient Average value Absolute max Max in NCC-RCC
C-1 4.5 12.9 15.7
C-2 2.1 9.7 9.7
C-3 3.2 9.9 14.8
C-4 2.1 22.9 22.9
C-5 1.5 5.3 9.9
C-6 1.2 8.4 8.4
C-7 2.3 9.4 12.7
C-8 1.4 4.6 5.1
C-9 1.7 4.9 10.8
C-10 2.6 14.0 14.0
C-11 2.5 15.4 15.4
C-12 2.2 14.0 14.0
C-13 2.9 15.3 15.3
C-14 4.7 24.5 24.5
Average + /− std 2.5 ± 1.1 12.2 ± 5.2 13.8 ± 5.2

In the first column, the average value found in all the areas of the implantation site below the coronary ostia; in the second column, the absolute maximum value on the entire implantation site; in the last column, the maximum stress in the region between the right and non-coronary leaflets, where the left bundle branch is located.