Skip to main content
. Author manuscript; available in PMC: 2020 Jun 19.
Published in final edited form as: IEEE Trans Med Imaging. 2017 Oct 10;37(2):504–515. doi: 10.1109/TMI.2017.2761756

Table I.

Mean Activity increases (%) of different Motion Corrections compared to NMC reconstruction in different organs of the 18F-FPDTBZ studies.

Subject Right kidney Left Kidney Pancreas
Gated NR-Discrete R-EBE NR-EBE Gated NR-Discrete R-EBE NR-EBE Gated NR-Discrete R-EBE NR-EBE
1 22.2 8.8 13.2 11.7 22.8 7.6 13.6 12.8 17.9 5.3 15.1 9.7
2 11.4 5.9 8.3 6.1 16.9 7.0 10.4 9.5 7.7 5.5 6.3 6.2
3 8.3 8.6 7.5 10.0 12.6 6.9 5.7 9.3 11.5 13.3 9.6 13.1
Mean 13.9 7.8 9.6 9.3 17.5 7.2 9.9 10.5 12.4 8.0 10.3 9.7