Skip to main content
. 2020 Jun 21;81(4):607–613. doi: 10.1016/j.jinf.2020.06.051

Table 4.

Measures of diagnostic accuracy of FebriDx MxA for identification of COVID-19, according to duration of illness.

n/n ≤7 days of
symptoms n = 129
Value (95% CI)
n/n >7 days of
symptoms n = 86
Value (95% CI)
Median (IQR) duration, days 4 (1 to 6) 14 (8 to 14)
Prevalence of COVID-19 52/129 40 (32 to 49) 53/86 62 (51 to 71)
Sensitivity 49/52 94 (84 to 98) 48/53 91 (80 to 96)
Specificity 65/77 84 (75 to 91) 29/33 88 (73 to 95)
Positive predictive value 49/61 80 (69 to 88) 48/52 92 (82 to 97)
Negative predictive value 65/68 96 (88 to 99) 29/34 85 (71 to 93)
Positive likelihood ratio 6.05 (3.58 to 10.0) 7.47 (2.96 to 19.0)
Negative likelihood ratio 0.07 (0.02 to 0.21) 0.11 (0.5 to 0.25)

IQR=Interquartile range, CI=Confidence interval.