Skip to main content
. 2019 Nov 7;109(7):802–809. doi: 10.1007/s00392-019-01567-y

Table 3.

Sensitivity, specificity, PPV, and NPV with 95% CI of the Microlife WatchBPO3 AFIB to detect paroxysmal AF in various settings

Sensitivity (%) Specificity (%) PPV (%) NPV (%)
Overall (24-h) AF detection 76.0 (74–78) 80.8 (79–83) 5.9 (5–7) 99.5 (99–100)
Day-time AF detection 68.4 (66–71) 79.3 (77–81) 5.1 (4–6) 99.4 (99–100)
Night-time AF detection 100 (99a–100) 84.9 (83–87) 8.8 (7–10) 100 (99a–100)
AF detection in < 80% successful BP readings 76.0 (59–96) 76.9 (70–79) 15.4 (8–21) 98.3 (97–100)
AF detection in > 80% successful BP readings 100 (99a-100) 83.1 (80–86) 0.5 (0–4) 100 (99a–100)

aOne-sided 97.5% confidence interval