Skip to main content
. 2020 Jun 24;58(7):e00128-20. doi: 10.1128/JCM.00128-20

TABLE 9.

Performance of the BioFire PN panel for AMR determinants compared to independent PCR/sequencing

Analytea Specimen Positive percent agreement
Negative percent agreement
TP/(TP+FN) % (95% CI) TN/(TN+FP) % (95% IC)
mecA/mecC + MREJ BAL 40/45 88.9 (76.5–95.2) 64/70 91.4 (82.5–96.0)
SPU 94/98 95.9 (90.0–98.4) 91/104 87.5 (79.8–92.5)
KPC BAL 2/2 100 (34.2–100) 148/149 99.3 (96.3–99.9)
SPU 7/7 100 (64.6–100) 284/284 100 (98.7–100)
NDM BAL 0/1 0 149/150 99.3 (96.3–99.9)
SPU 0/0 291/291 100 (98.7–100)
OXA-48 BAL 0/0 151/151 100 (97.5–100)
SPU 0/0 291/291 100 (98.7–100)
VIM BAL 0/0 151/151 100 (97.5–100)
SPU 1/1 100 (20.7–100) 289/290 99.7 (98.1–99.9)
IMP BAL 0/0 151/151 100 (97.5–100)
SPU 0/0 291/291 100 (98.7–100)
CTX-M BAL 6/7 85.7 (48.7–97.4) 144/144 100 (97.4–100)
SPU 8/10 80 (49.0–94.3) 280/281 99.6 (98.0–99.9)
a

Reported only when an applicable host organism is also detected by the BioFire PN panel (see Table 1, footnotes b, c, and d).