Skip to main content
. 2020 Jun 24;58(7):e01775-19. doi: 10.1128/JCM.01775-19

TABLE 2.

Clinical performance of the Revogene Strep A assaya

Revogene Strep A assay result Bacterial culture plus MALDI-TOF MS results (no.)
Positive Negative Total
Positive 151 24 175
Negative 3 426 429
Total 154 450 604
a

Sensitivity and specificity of the Revogene Strep A assay were 98.1 (95% CI, 94.4 to 99.3) and 94.7 (95% CI, 92.2 to 96.4), respectively, and the PPV and NPV were 86.3 (95% CI, 80.4 to 90.6) and 99.3 (95% CI, 98.0 to 99.8), respectively.