Skip to main content
. Author manuscript; available in PMC: 2020 Jun 25.
Published in final edited form as: Analyst. 2020 Mar 3;145(7):2525–2542. doi: 10.1039/c9an02250c

Table 3:

HemeChip diagnostic sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) in comparison to reference standard method.

SCD-SS vs. others SCD-SC vs. others SCD Trait vs. others Hb E Disease vs. others Hb E Trait vs. others
True positive, TP 33 14 129 15 57
True negative, TN 687 706 591 705 663
False Positive, FP 9 0 0 0 0
False negative, FN 0 0 0 0 0
Sensitivity, TP/(TP + FN) 100.0% 100.0% 100.0% 100.0% 100.0%
Specificity, TN/(TN + FP) 98.7% 100.0% 100.0% 100.0% 100.0%
PPV, TP/(TP + FP) 78.6% 100.0% 100.0% 100.0% 100.0%
NPV, TN/(TN + FN) 100.0% 100.0% 100.0% 100.0% 100.0%

HemeChip reports hemoglobin C/E/A2. Test location was used to differentiate co-migrating hemoglobins C and E.