Skip to main content
. 2020 Jun 26;19:54. doi: 10.1186/s12938-020-00794-z

Fig. 4.

Fig. 4

Non-target embolization (inaccuracy) for each catheter type and injection method. Catheter effect was nearly significant (p = 0.059), with a pairwise comparison of extremes (End-hole and Endobar) being highly significant (p =0.0032) and the ordering of outcome means (target and nontarget) across the three catheters being preserved. Injection method effect was not statistically significant across all catheter types (p = 0.1476); however, for the Endobar catheter, the difference in outcomes was significant in a pairwise comparison between the Endobar manifold and manual injections (p = 0.041)