. 2020 Jun 26;19:54. doi: 10.1186/s12938-020-00794-z

Table 1.

Data for each test conducted

Catheter Injection method Injection ratea (mL/min) Post-injection flow rates per vessel, vessels 1–12 (mL/min) Number non-target vessels embolized Percent microspheres to target vessels Total microspheres counted
End-hole Syringe pump 1.50 2.06 5 24.48% 192
Syringe pump 1.50 1.45 6 52.73% 1117
Pressure replicated 3.40 1.27 6 52.06% 559
Pressure replicated 3.40 0.33 8 30.19% 2183
Manual 8.29 0.39 7 38.60% 3927
Manual 10.80 0.22 8 33.35% 3094
Surefire Syringe pump 1.50 1.42 6 77.91% 2218
Syringe pump 1.50 2.16 3 73.80% 1500
Pressure replicated 3.40 2.04 4 62.10% 1103
Pressure replicated 3.40 1.32 6 63.53% 2328
Manual 10.75 1.43 6 47.31% 2579
Manual 7.58 1.21 6 59.58% 3105
Endobar Endobar manifold 1.50 3.63 0 96.30% 729
Endobar manifold 1.50 3.57 0 94.36% 780
Syringe pump 1.50 2.91 2 81.71% 257
Syringe pump 1.50 1.23 0 99.43% 174
Pressure replicated 3.40 2.04 4 85.06% 1486
Pressure replicated 3.40 1.35 6 65.43% 2183
Manual 5.25 0.26 8 37.98% 2383
Manual 6.56 4.94 0 99.67% 2148
Manual  5.62 1.85 5 67.96% 1910

aManual injection rates were estimated from video recorded during each test