. 2020 Jun 26;10:10425. doi: 10.1038/s41598-020-66829-6

Table 1.

Accuracy and precision results within-day and between-days for the qualification of HPRT.siRNA antisense POE immunoassay in Sprague-Dawley serum and liver tissue homogenate.

Qualification parameters of HPRT-siRNA POE immunoassay in Sprague-Dawley serum (antisense strand) Qualification parameters of HPRT-siRNA POE immunoassay in Sprague-Dawley liver homogenate (antisense strand)
Within-day (n = 18) Between-day (n = 18) Within-day (n = 18) Between-day (n = 18)
%CV %Bias %CV %Bias %CV %Bias %CV %Bias
LQC 7.0 to 13.8 −20.2 to 4.0 16.1 −5.8 LQC 3.0 to 19.6 −8.2 to 8.4 11.6 −2.1
MQC 1.0 to 19.3 −11.5 to 1.3 18.7 −4.6 MQC 6.0 to 32.8 −11.3 to 16.9 16.4 0.9
HQC 4.1 to 24.2 −18.4 to 24.8 25.7 0 HQC 12.0 to 29.6 −34.4 to 24.0 26.7 −3.2