Skip to main content
. 2020 Jun 19;2020:7625725. doi: 10.1155/2020/7625725

Table 2.

The refractive error of the targeted refraction using the Barrett Universal II and the SRK/T formulas.

Keratometry Number of eyes Refractive error Barrett Universal II SRK/T p value
Entire 335 Prediction error (D) 0.04 ± 0.42 (−1.37 to 1.32) −0.10 ± 0.53 (−2.75 to 1.19) <0.001
Absolute error (D) 0.33 ± 0.27 (0.00 to 1.37) 0.38 ± 0.38 (0.00 to 2.75) 0.006
Median absolute error (D) 0.25 0.30

Flat K group 69 Prediction error (D) 0.15 ± 0.42 (−0.65 to 1.32) 0.27 ± 0.40 (−0.63 to 1.19) <0.001
Absolute error (D) 0.35 ± 0.27 (0.01 to 1.32) 0.37 ± 0.30 (0.00 to 1.19) 0.334
Median absolute error (D) 0.29 0.32

Normal K group 180 Prediction error (D) −0.02 ± 0.39 (−1.08 to 1.06) −0.05 ± 0.39 (−1.15 to 1.09) 0.138
Absolute error (D) 0.29 ± 0.24 (0.00 to 1.08) 0.30 ± 0.25 (0.00 to 1.08) 0.925
Median absolute error (D) 0.23 0.23

Steep K group 86 Prediction error (D) 0.08 ± 0.49 (−1.37 to 1.26) −0.49 ± 0.63 (−2.75 to 0.66) <0.001
Absolute error (D) 0.38 ± 0.31 (0.01 to 1.37) 0.57 ± 0.56 (0.00 to 2.75) 0.003
Median absolute error (D) 0.27 0.37

K = keratometric readings; D = diopter.