Skip to main content
. Author manuscript; available in PMC: 2020 Aug 1.
Published in final edited form as: Magn Reson Med. 2020 Jan 21;84(2):928–948. doi: 10.1002/mrm.28167

TABLE 3.

Comparison of in vivo T1 measurements between Multitasking DCE and MOLLI

Multitasking (ms) MOLLI (ms) ICC Paired t-test (p-value)
Pre-contrast Control 757 ± 55 818 ± 46 0.820 .024
Tumor 1571 ± 210 1621 ± 171 0.904 .180
Non-tumor 1183 ± 338 1245 ± 352 0.973 .428
Post-contrast Control 419 ± 74 470 ± 80 0.924 <.001
Tumor 386 ± 63 426 ± 56 0.958 .013
Non-tumor 417 ± 58 460 ± 49 0.922 .043

Note: The mean values and SD of pre-contrast and post-contrast T1 from Multitasking DCE and MOLLI are listed for normal pancreas in control group, tumor, and nontumoral regions in patients. The ICC and p-value by paired t-test between T1 measured by MOLLI and Multitasking are provided. The ICC of each category indicates good agreement of in vivo T1 measurement between Multitasking DCE and MOLLI.