Skip to main content
. 2020 May 12;4:7. doi: 10.1186/s41824-020-00076-w

Table 2.

CR relative percentage difference between no-array and the anterior (A) part of each array for each sphere size. The CR was estimated for both reconstruction methods (MRAC and off-line DCTAC)

Sphere size (mm) 10 13 17 22 28 37
MRAC PET/MRI 32ch (%) 4.8 2.8 5.2 3.8 2.1 − 1.5
mMR12ch (%) 15.1 2.6 − 4.0 − 5.4 2.0 5.2
MRI 32ch (%) − 20.2 − 17.9 − 22.7 − 16.4 − 13.1 − 17.3
p value 0.003 0.01 0.00 0.001 0.00 0.00
DCTAC PET/MRI 32ch (%) 3.8 0.9 3.6 2.0 3.7 2.4
mMR12ch (%) 6.0 8.0 4.1 3.5 4.2 6.6
MRI 32ch (%) − 10.6 − 7.7 − 9.5 − 11.1 − 6.8 − 6.2
p value 0.007 0.011 0.129 0.053 0.050 0.001