Table 4.

Diagnostic performances of a GEL technique compared to a TUBE technique in feline AB system blood typing of 143 feline whole-blood samples anticoagulated with EDTA, where GEL mixed pattern results were considered as negative results.

Statistics Blood type with GEL (considering mixed-field pattern as negative
results) compared to TUBE technique
A B AB
Value 95% CI Value 95% CI Value 96% CI
(%) (%) (%)
Sensitivity (Se) 100.00 96.31–100.00 96.00 79.65–99.90 95.00 75.13–99.87
Specificity (Sp) 100.00 92.13–100.00 99.15 95.37–99.98 99.19 95.55–99.98
Positive likelihood ratio (PLR) 113.28 16.06–798.86 116.85 16.55–825.09
Negative likelihood ratio (NLR) 0.00 0.04 0.01–0.28 0.05 0.01–0.34
Positive predictive value(PPV) 100.00 89.65 55.11–98.39 71.48 26.20–94.65
Negative predictive value (NPV) 100.00 99.69 97.94–99.95 99.89 99.27–99.98
Accuracy 100.00 97.45–100.00 98.93 95.55–99.92 99.10 95.83–99.95