Table 1. Description of the most used RT-PCR for diagnosing SARS-Cov-2.

  • Institution/ country

  • Primer ID

  • Method

  • Findings

  • China’s CDC – China29,56

  • ORF1ab, E and N

  • Three singleplex RT-PCR using fluorescent-labeled probes

  • Lowest detection limit of ORF1ab was 203 copies/mL

  • Lowest detection limit of E gene was 664 copies/mL

  • Lowest detection limit of N gene was 667 copies/mL

  • Institut Pasteur – France29

  • RdRp gene/nCoV_IP2

  • RdRp gene/nCoV_IP4

  • Two singleplex RT-PCR or one multiplex RT-PCR with fluorescent-labeled probes

  • 95% assertiveness in samples containing ~

  • 100 viral copies

  • The multiplex RT-PCR detects 10 viral copies

  • No cross-reactivity with other viruses

  • CDC – USA29,57

  • 2019-nCoV_N1

  • 2019-nCoV_N2

  • 2019-nCoV_N3

  • Three singleplex RT-PCR using fluorescent-labeled probes

  • 95% assertiveness in samples containing

  • 100,5 RNA copies/µL using automatic RNA extraction and 100

  • RNA copies/µL using manual RNA extraction

  • Charité – Germany29,58

  • RdRP_SARSr

  • E_Sarbeco

  • Two singleplex RT-PCR, the screening made by the E gene amplification confirmed by RdRP

  • Lowest detection limit of RdRp and E_Sarbeco 3.8 and 5.2 copies/reaction, respectively

  • HKU - Hong Kong29,59

  • ORF1b-nsp14

  • N

  • Two RT-PCR, firstly the N gene amplification with fluorescent-labed probe followed by confirmation by amplifying the ORF1b-nsp14 with fluorescent-labeled probe

  • N gene Ct values up to 35.43 were positive corresponding to 15 copies/reaction

  • ORF1b Ct values up to 38.97 were positive corresponding to 1.5 copies/reaction

  • National Institute of Health – Thailand29,59

  • N

  • Singleplex RT-PCR with fluorescent-labeled probe

  • N gene Ct values were positive up to 38.12, corresponding to 15 copies/reaction