Skip to main content
. 2020 Jun 30;20:612. doi: 10.1186/s12885-020-07077-9

Table 2.

LoD and LoQ of the clonoSEQ Assay by MRD cell counts and MRD frequency

Measure Malignant cellsa (95% CI) 500 ng DNA input frequency (95% CI) 20 μg DNA input frequency (95% CI)
LoD 1.903 (1.75–2.07)

2.26 × 10− 5

(2.01 × 10− 5–2.53 × 10− 5)

6.77 × 10− 7

(6.02 × 10− 7–7.61 × 10− 7)

LoQ 2.390 (1.90–9.14)

2.39 × 10−5

(2.26 × 10− 5–7.01 × 10− 5)

1.76 × 10−6

(6.77 × 10−7–4.09 × 10− 6)

CI confidence interval, LoD limit of detection, LoQ limit of quantitation, MRD minimal residual disease

aCalculated from samples with 500 ng and 20 μg of DNA input