. 2020 Jun 30;20:612. doi: 10.1186/s12885-020-07077-9

Table 3.

Summary of the clonoSEQ Assay precision

%CV attributed to each variable at cell inputsa
Lot-to-lot variability Number of input cancer cells 2.14 6.13 21.44 61.26 214.4 612.56
Instrument set (%) 0 1 0 1 1 1
Operator (%) 2 0 1 2 0 0
Processing day (%) 0 0 1 1 0 3
Processing run (%) 0 0 1 0 0 0
Reagent lot (%) 0 0 0 1 2 1
Residual variability (%) 68 49 28 23 19 18
Total MRD measurements, n 3456 3456 3564 3960 3960 3828

%CV percent coefficient of variance, ALL acute lymphoblastic leukemia, CLL chronic lymphocytic leukemia, MM multiple myeloma

aThese values were aggregated across diseases (ALL, CLL, and MM) and total DNA input levels