table I.
Item name | Description | References |
---|---|---|
Code | PENELOPE (2006) | 13 |
Validation | The Monte Carlo code was previously validated for full rotation CT (for both axial and helical scans, see references). In the present study, the localizer validation is described in Section 2.B | 10,11 |
Source Description | Particle type: x-ray Prebowtie: The source is modeled as a point source with an effective beam width to account for the penumbra of the beam. Prebowtie spectra are provided by the manufacturer in 0.5-keV bin increments for a 120-kVp tube potential. Postbowtie: The prebowtie spectrum is filtered by the bowtie using a method described in Section 2.A of the present paper, originally described by Li et al. (2014) |
14,15 |
Cross sections | Cross sections are generated using the default material.f provided with PENELOPE, which has a large database of predefined common materials | |
Transport parameters | Transport of secondary electrons is not included | |
VRT or AEIT | Variance reduction was performed using the particle-splitting technique | 10 |
Scored Quantities | The Monte Carlo scored quantity is Kerma (eV) in each organ per number of starting (prebowtie) photons | |
Number of histories | 8.00E+07 | |
Postprocessing | The postprocessing for the present study is described in Section 2.A |