Table VII.

The fitting parameters (a, b), R2, and root-mean-square from the residual (RMSE), for PA localizer acquisition geometry across the population of XCAT phantoms (n = 57).

Organ PA
a b R2 RMSE
Adrenal 1.377 −0.067 0.670 0.022
Bladder 0.025 −0.076 0.239 0.001
Bones 0.398 −0.034 0.622 0.010
Brain 0.004 0.009 0.023 0.001
Breast 0.897 −0.099 0.849 0.005
Esophagus 0.446 −0.038 0.398 0.017
Eye 0.001 0.020 0.050 0.001
Gall bladder 1.600 −0.109 0.678 0.011
Heart 0.622 −0.059 0.628 0.012
Kidney 2.629 −0.090 0.766 0.022
Large intestine 0.313 −0.077 0.534 0.006
Larynx 0.102 −0.011 0.014 0.017
Liver 0.468 −0.051 0.427 0.015
Lung 0.540 −0.036 0.602 0.014
Marrow 0.284 −0.037 0.656 0.006
Ovaries 0.018 −0.041 0.514 0.014
Pancreas 0.971 −0.076 0.661 0.013
Prostate 0.018 −0.041 0.514 0.014
Skin 0.114 −0.032 0.649 0.003
Small intestine 0.693 −0.091 0.632 0.008
Spleen 0.509 −0.042 0.257 0.026
Stomach 0.448 −0.055 0.411 0.014
Testes 0.017 −0.046 0.514 0.014
Thymus 0.384 −0.051 0.407 0.013
Thyroid 0.216 −0.031 0.240 0.012
Trachea 0.457 −0.037 0.514 0.014
Uterus 0.017 −0.046 0.514 0.014
Vagina 0.007 −0.046 0.514 0.014