Intraday and interday assay precision and accuracy for fluconazole in serum samples

Conditions Concentration (μg/mL) Accuracy (%) Precision (%)
Intraday 0.01 −4.6 3.1
0.1 4.7 4.6
1.0 −0.4 2.1
10.0 7.4 3.4
50.0 5.0 6.0
100.0 −7.7 5.8
Interday 0.01 −4.4 4.3
0.1 0.8 2.7
1.0 3.0 0.3
10.0 3.2 2.5
50.0 −2.5 2.5
100.0 −0.3 5.4