. 2020 Jun 1;10(16):7211–7230. doi: 10.7150/thno.47281

Table 2.

PCI and MRI binary tissue damage volume centroid comparison

Rabbit # 150% psub Total (mm) 150% psub, Acoustic Axis (mm) 150% psub, Axial Plane (mm) 100% psub, Total (mm) 100% psub Acoustic Axis (mm) 100% psub Axial Plane (mm)
1 0.5 0.1 0.5 0.4 0.1 0.4
2 0.3 0.1 0.3 0.4 0.1 0.4
3 0.6 0.2 0.6 0.5 0.2 0.5
4 1.4 1.3 0.4 1.9 1.7 0.8
5 0.2 0.1 0.2 0.4 0.1 0.4
6 1.7 1.6 0.4 1.5 1.4 0.5
7 1.4 1.3 0.2 1.0 0.2 1.0
8 0.4 0.2 0.4 NA NA NA
9 1.7 1.6 0.4 NA NA NA

Euclidian distance between the centroids of the tissue damage volumes predicted via sonication-aggregate 3D PCI and those measured using T2*w MRI. Total distances are provided along with the components along the transducer's acoustic axis (superior-inferior direction) and within the orthogonal axial plane. Data are provided on an animal-wise basis (cohort #1) and are stratified based on the exposure level.