Skip to main content
. 2020 Jun;52(2):118–125. doi: 10.1182/ject-1900037

Table 3.

Efficacy of air bubble removal based on counts at a flow rate of 75 mL/h.

Filter Model Percentage Removed (%)* 95% Confidence Interval
TCBINF044G 86.2 ± 21.1†‡ 64.1–108.3
REF473036 100.0 ± .0† 100.0–100.0
Control −.8 ± 1.5 −2.7–1.0

*Results denote means ± SD.

p < .0001 compared with the control group.

p > .05 compared with the REF473036 filter model group.