Table 2|.
Troubleshooting tablea
Step | Problem | Possible Reason | Solution |
---|---|---|---|
11 | [18F]Fluoride is retained on QMA cartridge | Connections to QMA cartridge are leaking or wrong cartridge used | Check connections, replace as needed |
15 | Low incorporation of 18F | Equipment malfunction, decomposition of reactants | Ensure synthesis module is clean and functional, check and replace reactants as needed |
15 | BPin precursor 1 ceases to be commercially available | Economic / patent considerations | Prepare Bpin precursor 1 according to the procedure provided in Box 2 |
16 | Low purity after deprotection | Ascorbic acid is degrading | If it becomes an issue, check and replace ascorbic acid if needed. |
18 | Slow (or no) loading of crude reaction onto HPLC loop | HPLC loop is blocked | Clean HPLC loop |
18 | More radioactive impurities in HPLC trace | Not enough acid in Deprotection Solution | Increase the amount of 36.5 – 38.0% HCl in Deprotecting Solution by 0.1 mL increments, or use fresh 36.5 – 38.0% HCl. |
19, 27 | Low chemical and/or radiochemical purity | [18F]FDOPA was not adequately purified | Clean or replace guard column and semi-preparative HPLC column |
20,21, 30 | Too much MeCN in [18F]FDOPA product | MeCN used during reformulation was not adequately removed | Optimize rinsing and drying in Steps 20 and 21 |
Many of the problems in this table can be be avoided by following cGMP.