Skip to main content
. Author manuscript; available in PMC: 2020 Nov 1.
Published in final edited form as: Nat Protoc. 2020 Apr 8;15(5):1742–1759. doi: 10.1038/s41596-020-0305-9

Table 2|.

Troubleshooting tablea

Step Problem Possible Reason Solution
11 [18F]Fluoride is retained on QMA cartridge Connections to QMA cartridge are leaking or wrong cartridge used Check connections, replace as needed
15 Low incorporation of 18F Equipment malfunction, decomposition of reactants Ensure synthesis module is clean and functional, check and replace reactants as needed
15 BPin precursor 1 ceases to be commercially available Economic / patent considerations Prepare Bpin precursor 1 according to the procedure provided in Box 2
16 Low purity after deprotection Ascorbic acid is degrading If it becomes an issue, check and replace ascorbic acid if needed.
18 Slow (or no) loading of crude reaction onto HPLC loop HPLC loop is blocked Clean HPLC loop
18 More radioactive impurities in HPLC trace Not enough acid in Deprotection Solution Increase the amount of 36.5 – 38.0% HCl in Deprotecting Solution by 0.1 mL increments, or use fresh 36.5 – 38.0% HCl.
19, 27 Low chemical and/or radiochemical purity [18F]FDOPA was not adequately purified Clean or replace guard column and semi-preparative HPLC column
20,21, 30 Too much MeCN in [18F]FDOPA product MeCN used during reformulation was not adequately removed Optimize rinsing and drying in Steps 20 and 21
a

Many of the problems in this table can be be avoided by following cGMP.