Table 3.
Patient details | Device and lead number | Indication | Dwell time of oldest lead (months) | Equipment used | Complication and treatment | Outcomes |
---|---|---|---|---|---|---|
80 years, male | Dual chamber pacemaker | Infection | 124 | Locking stylet, simple sheath | Cardiac avulsion | Full recovery after surgical repair |
83 years, female | Single chamber pacemaker | Infection | 151 | Locking stylet, simple sheath | Pericardial effusion, percutaneous drainage | Full recovery |
37 years, female | Dual chamber pacemaker with abandoned leads (4 leads total) | Infection, local | 228 | Locking stylets, femoral snare (needle eye) | Pericardial effusion post-procedure, percutaneous drainage | Full recovery |
77 years, male | Dual chamber pacemaker | Redundant lead | 178 | Locking stylet, Evolution (13F), femoral snare (needle eye) | Cardiac avulsion during use of femoral snare | Surgical repair, but died within 30 days |
65 years, female | CRT-D | Lead failure | 141 | Locking stylet, simple sheath, Evolution | Right ventricular injury | Full recovery after surgical repair |
46 years, male | Dual chamber pacemaker | Systemic infection | 47 | Locking stylet | Pericardial tamponade | Full recovery after surgical repair |
83 years, female | Dual chamber pacemaker with abandoned leads (total 4 leads) | Infection | >120 | Evolution, femoral snare | Pericardial effusion | Uncomplicated recovery after percutaneous drainage |
88 years, female | Dual chamber pacemaker (2 leads) | Perforated lead, lying in pleural space | 2 | Locking stylet | Oesophageal injury from transoesophageal echo probe | Recovered after long hospital stay |
80 years, female | Dual chamber pacemaker (2 leads) | Infection | 285 | Locking stylet, compression coil, Evolution RL (11F) | Peri-procedural stroke | Recovered |
43 years, male | Dual chamber ICD with abandoned leads (3 leads) | Infection | 66 | Evolution | Pericardial tamponade | Full recovery after surgical repair |
47 years, female | Single chamber ICD | Lead malfunction | 59 | Evolution RL | Right pleural collection of blood drained percutaneously at 1 day after extracting right sided device | Full recovery. No surgery, no transfusion required |
78 years, female | Dual chamber PPM | Local infection | Unknown | Locking stylet, compression coil, Evolution RL (11F) | Injury to right ventricle | Full recovery after surgical repair |
75 years, female | CRT-P (including Starfix lead) | Lead malfunction | 15 | Locking stylet, compression coil, Evolution RL (9F) | Injury to coronary sinus | Full recovery after surgical repair |
63 years, male | CRT-D | Lead malfunction | 83 | Locking stylet, compression coil, Bulldog, Evolution RL (13F) | Injury to inferior vena cava causing pericardial tamponade | Full recovery after surgical repair |
74 years, male | Dual chamber PPM, abandoned leads (total 4 leads) | Venous occlusion, redundant leads, need for upgrade | 356 | Locking stylets, Evolution 9F | Pericardial tamponade due to RV perforation | Full recovery after surgical repair |
78 years, female | Dual chamber pacemaker | Local infection | Unknown | Liberator, compression coil, Evolution 9F | Tamponade due to RV perforation | Full recovery after surgical repair |
79 years, male | CRT-D with redundant leads (total 7 leads) | Septicaemia | 79 | Liberator, compression coil, Evolution RL 13F | Septic shock | Peri-procedural death |
37 years, male | Dual chamber ICD | Lead malfunction | 83 | Simple stylet | Pericardial tamponade | Full recovery after surgical repair |
83 years, female | CRT-D | Local infection | 126 | Liberator, compression coil, Evolution RL 13F | Pericardial tamponade | Initial recovery after surgical repair, but died at 30 days |
65 years, male | CRT-D | Systemic infection | 105 | Liberator, compression coil, Evolution RL 13F | Left-sided haemothorax | Full recovery after surgical placement of a chest drain |
61 years, female | Single chamber pacemaker with redundant leads (3 leads in total) | Septicaemia | 208 | Liberator, compression coil, Evolution RL 13F | Injury to junction of SVC and RA leading to pericardial tamponade | Died at 2 days post-procedure despite surgical repair |
48 years, male | Single chamber ICD | Systemic infection | 4 | Liberator, compression coil, Evolution RL 13F | Damage to tricuspid valve requiring subsequent valve surgery | Full recovery after surgical repair |
CRT-D, cardiac resynchronization therapy with defibrillator; CRT-P, cardiac resynchronization therapy pacemaker; ICD, implantable cardioverter-defibrillator; PPM, permanent pacemaker; RA, right atrium; RV, right ventricle.