Abstract
The tumor-involved regions of head and neck cancer (HNC) have complex anatomical structures and vital physiological roles. As a consequence, there is a need for high sensitivity and high spatial resolution dedicated HNC PET scanner. The purpose of this study is to evaluate and optimize system design that includes detecting materials and geometries. For the detecting material, two scanners with the same two-panel geometry based on CZT and LYSO were evaluated. For the system geometry, four CZT scanners with two-panel, lengthened two-panel, four-panel, and full-ring geometries were evaluated. A cylinder phantom with sphere lesions and an XCAT phantom in the head and neck region were simulated. The results showed that the sensitivity of the 40-mm thickness CZT system and the 20-mm thickness LYSO system were comparable. However, the multiple interaction photon events recovery accuracy of the CZT system was about 20% higher. The in-panel and orthogonal-panel spatial resolutions of CZT are 0.58 and 0.74 mm, while those of LYSO are 0.70 and 1.40 mm. For system geometry, the four-panel and full-ring scanners have a higher contrast recovery coefficient (CRC) and contrast-to-noise ratio (CNR) than the two-panel and lengthened two-panel scanners. However, a 5-mm lesion in the XCAT phantom was visualized within 6 min in the two-panel system.
Index Terms—: Head and neck cancer, Dedicated PET, System design, Monte Carlo simulation
I. Introduction
Head and neck cancer (HNC) accounts for approximately 4% of all cancers in the United States [1] and the overall annual mortality rate is 23% [2]. Whole-body positron emission tomography (PET) and its combination with computed tomography (CT) are commonly used for HNC diagnosis, staging, treatment planning, and assessing response to therapy [3]–[6]. Compared to CT, magnetic resonance imaging, sonographic and histopathological findings, PET imaging shows the highest sensitivity and specificity for detecting lymph node metastases of HNC [7].
However, lesions in this region can be challenging to diagnose due to the thin, soft tissues within the neck, which require a high-resolution imaging system. The spatial resolution of whole-body PET is typically 4 to 6 mm [8], [9]. For structures less than twice the reconstructed image resolution, the true amount of activity is not completely captured [10]. The poor spatial resolution of whole-body PET hinders the precise delineation of the primary tumor of HNC and limits the detection of tumor involvement in lymph nodes smaller than 4 to 5 mm. Besides, it also results in a large number of reported false-negatives in lymph nodes (for example 80%) [11]–[13]. Researchers [14] from Duke University Medical Center have introduced a dedicated HNC PET acquisition protocol with longer scan time in the HN bed position to improve the detection ability of PET imaging in HNC. The dedicated HN protocol has advantages in detecting lymph nodes smaller than 15 mm compared to the standard protocol. However, in terms of evaluating primary tumors and detecting lymph nodes smaller than 10 mm, there are no significant differences between the standard protocol and the dedicated HN protocol [13], [14].
Due to the complex anatomy and vital physiological role of the tumor-involved structures, the goal of HNC treatment is not only to improve survival outcomes but also to preserve organ functions [15]. An improvement in resolution in a PET image to better define the boundary of tumors is significant for the treatment planning and monitoring of HNC. In supraglottic squamous cell carcinoma, for example, if it involves thyroid cartilage, it is T4 and unresectable. If it does not involve thyroid cartilage, it is T3 and can be cured with surgery. Another example is that if supraglottic squamous cell carcinoma does not cross the anterior commissure, patients can have supraglottic laryngectomy and can be cured without losing voice.
As a result, there is a need for high spatial resolution and high sensitivity PET imaging in HNC. We are designing a high-resolution add-on dedicated HNC PET scanner to complement the whole-body PET scanner. This system will image the patients right after the whole-body PET scanning without injecting any extra dose to the patient. The dedicated system will provide extra information for scenarios when radiologists are looking for small lymph nodes or well-defined tumor boundaries. FIG. 1 shows an illustration of the add-on dedicated system [16], [17].
Fig. 1.

Illustration of an add-on dedicated head and neck cancer PET scanner integrated into the standard whole-body PET/CT imaging workflow. The gantry is implemented to interface with the patient bed to image the patient right after the whole-body PET scanning.
With more accurate detection of small lymph nodes and estimation of the extent of tumor growth, the dedicated head and neck will provide physicians with more freedom to choose treatment options including surgery, radiation therapy, and chemotherapy. It helps with more accurate radiation dose planning and will lead to better patient outcomes such as preserving organ functions. It also improves confidence in differentiating post-treatment changes from tumor recurrence.
To design such a dedicated system, the first consideration is the detecting material. Cadmium zinc telluride (CZT) detectors and lutetium-yttrium oxyorthosilicate (LYSO) combined with silicon photomultiplier (SiPM) detectors are investigated for HNC PET scanner. Compared with LYSO, CZT has a high energy resolution, sub-millimeter intrinsic spatial resolution [18], [19] and intrinsically achievable depth-of-interaction (DOI) information. It has been used for the small field of view (FOV) and high-resolution PET applications such as small animal imaging [20]–[23]. The high energy resolution and high spatial resolution of CZT detectors are also important for correctly identifying the first interaction position in multiple interaction photon events (MIPEs) using Compton kinematics.
Another important consideration for designing a dedicated PET is the system geometry. Compared with a whole-body PET scanner, dedicated PET often utilizes smaller and compact geometry to adapt to the dedicated imaging environment and improve the system sensitivity [24]. For example, dedicated brain PET has ring and helmet geometries [25]–[28], and dedicated breast PET has two-panel and ring geometries [29]–[32]. In this work, we consider two-panel, lengthened two-panel, four-panel, and full-ring geometries for the sake of high sensitivity and patient’s comfort.
The purpose of this study is to evaluate and optimize system design among different detecting materials and geometries. For detecting material, two scanners with the same two-panel geometry based on CZT and LYSO are compared in terms of photon coincidence sensitivity, MIPEs recoverability, noise equivalent count (NEC) rate, and spatial resolution. For system geometry, CZT scanners with two-panel, lengthened two-panel, four-panel, and full-ring geometries are compared in terms of contrast recovery coefficient (CRC) and contrast-to-noise ratio (CNR) of reconstructed images.
II. Method and materials
A. Detecting material
Two systems with the same stationary two-panel geometry based on CZT and LYSO were built in GATE [33]. Based on the human head size [34], the panel was set as 150×200 mm2, and the distance between the two panels was 200 mm. For the LYSO system, the LYSO crystal segment size was 1×1 mm2 and the crystal thickness was 20 mm. The fill factor was 86.5%. The energy resolution, time resolution, and DOI resolution were assumed to be 15%, 400 ps, and 2 mm, respectively [35]–[37]. The time window and energy window for selecting coincidence events were 1 ns and [400, 620] keV, respectively. For the CZT system, the crystal size was 40×40×5 mm3, and the energy resolution, time resolution, and intrinsic spatial resolution were set to 2%, 8 ns, and 1×1×1 mm3, respectively [19]. The time window was 15 ns and energy window was [490, 530] keV. The deadtime of the LYSO scanner and CZT scanner were set as 1 μs and 10 μs, based on PETsys TOFPET2 ASIC [38] and Kromek RENA3 ASIC [39], respectively.
The photon coincidence sensitivity was defined as the coincidence rate divided by the point source activity. Generally, only photons without scattering were used to constitute coincidences (P-P coincidence). To further improve sensitivity, MIPEs have been recovered with different methods [40]–[43]. In this paper, Compton kinematics [40] was used to recover coincidences (P-CP coincidence) between annihilation photons that had a photoelectric event (P photon) and photons that had a Compton event before the photoelectric event (CP photon). In Compton kinematics, the scattering angle can be computed by energy,
| (1) |
where Ei is the incident photon energy, Es is the scattered photon energy, and m0c2 is the rest mass of an electron. For PET applications, Ei = m0c2 = 511 keV. The scattering angle can also be computed using the interaction position information,
| (2) |
where and are the directions of incident and scattered photons. FIG. 2 shows the principle to identify the Compton event in a P-CP coincidence. To compare the sensitivity and MIPE recoverability of the two systems and check the performance of Compton kinematics under a high single rate, 10 point sources with different activities (7, 14, …, 70 MBq) placed at the FOV center were simulated separately. Since the crystal size of many previously published dedicated brain PET scanners are around 2 mm [25], [27], [44]–[46], we compared the Compton recovery accuracy of systems based on 1-mm crystal size and 2-mm crystal size respectively.
Fig. 2.

Using Compton kinematics to identify the Compton event in a P-CP coincidence. The sequence with a smaller |θE − θp| is picked up.
NEC rate incorporates the noise effects of random and scatter counts and is an indicator of the signal-to-noise ratio for PET systems [47].
| (3) |
where T, S, and R are true, random and scatter coincidence rates respectively. To consider the effect of activity outside of the FOV, the phantom used for the NEC study contained three components, which represented the brain, neck, and torso. The brain component was a 130-mm diameter and 80-mm height cylinder with a 45.6 kBq/cm3 concentration activity [48], and the torso component was a 260-mm diameter and 200-mm height cylinder with 5.7 kBq/cm3. Both the brain and the torso were placed outside the FOV. For the neck component [14] (cylinder, 110-mm diameter, 126-mm height), 10 different concentration activity (5.7, 11.4, …, 57 kBq/cm3) were investigated.
For a stationary two-panel geometry, the incomplete angular sampling would cause the orthogonal-panel spatial resolution to be worse than the in-panel spatial resolution. To understand the difference, a point source was placed at the FOV center, and the line profiles along the orthogonal-panel and in-panel directions in the reconstructed image were fitted separately to measure the spatial resolution. Images were reconstructed with a list-mode 3D maximum likelihood expectation maximization (MLEM) algorithm [49] through the gpurecon program [50]. Time-of-flight (TOF) is known to have the large potential for image quality improvement and more accurate quantification (signal-to-noise ratio) for a given number of counts [51], [52], and TOF was incorporated in the image reconstruction of the LYSO system. As a comparison, the CZT system did not utilize TOF.
B. System geometry
For the system geometry study, four CZT systems with two-panel, lengthened two-panel, four-panel, and full-ring geometries were compared, as is shown in FIG. 3. All four systems have 150-mm thickness in the z axis. The panel size of the two-panel and the four-panel systems was 40×200 mm2, while the panel in the lengthened two-panel system was 40×320 mm2. The average human head size was 145 mm (head breadth) × 194 mm (head thickness) in the United States, and the maximum head size was 174 mm × 239 mm [34]. Since the distance between detector panels was adjustable, the distance between panels was set as 24 cm based on the average human head size. On the contrary, the full-ring structure did not have the flexibility to adjust its geometry, so the inner diameter was set as 30 cm based on the maximum human head size. The time resolution and energy resolution were set as 8 ns and 2%, respectively. The time window was 15 ns and energy window was [490, 530] keV. P-CP coincidences were used in image reconstruction.
Fig. 3.

Four CZT systems with two-panel (top-left), lengthened two-panel (top-right), four-panel (bottom-left), and full-ring (bottom-right) geometries were used for the phantom study. The x and y axes are shown in the figure, and the z axis is perpendicular to the paper. All four systems have the same 150-mm thickness in the z axis.
For the cylinder phantom study, four hot spheres with diameters 3, 4, 6 and 8 mm were placed in a 126-mm height, 110-mm diameter water phantom. The background concentration activity was 5.7 kBq/cm3, and the hot-to-background ratio was 8:1 [14]. We further compared the four geometries based on 1-mm crystal size and 2-mm crystal size. Each system had a 2-min data acquisition.
The same MLEM gpurecon program without TOF was used for image reconstruction. For the 1-mm crystal size, the voxel size of reconstructed image was 0.5×0.5×0.5 mm3. As a comparison, the voxel size of the 2-mm crystal size was 1×1×1 mm3. The image reconstruction had 10 iterations, which was chosen to maximize the CNR of the 3-mm sphere in the cylinder phantom study. Data corrections for scatter coincidence and random coincidence were not applied. No regularization or a post-reconstruction filter was used.
Image quality was evaluated based on CRC and CNR [53],
| (4) |
where Chot and Cbkg are the average voxel value in a hot sphere and background region of interest (ROI), respectively, ahot and abkg are the ground-truth concentration activity, and σbkg is the standard deviation of the voxel values in the background ROI.
For the XCAT phantom [54] study, simulations of the lengthened two-panel system and the four-panel system on the XCAT phantom in head and neck region (200×240×150 mm3 in x, y, and z axes) were performed. The 18F-FDG concentration activity in normal tissue, spinal cord, salivary gland, and brain were 3.4, 7.1, 7.8, and 16.9 kBq/cm3 based on clinical study, respectively [14], [55]–[57]. A sphere tumor with 5 mm diameter was put inside the phantom and the tumor concentration activity was 27.2 kBq/cm3. Each system had a 6-min data acquisition [14].
C. Comparison with a whole-body scanner
To validate the benefits of the add-on dedicated system, another simulation was performed to compare the two-panel scanner with one whole-body PET scanner (GE Discovery MI 4-ring PET scanner). The same cylinder phantom (11-cm diameter and 12.6-cm length) were simulated with both systems, respectively. Hot spheres (nine 3-mm diameter, nine 4-mm diameter, five 6-mm diameter, five 8-mm diameter) were placed in the central slice of the axial direction of the phantom. The background activity was 5700 Bq/cm3, and the sphere to background ratio was 8:1.
III. Results
A. Detecting material
The sensitivity of the CZT and LYSO systems under different source activities is shown in FIG. 4. At 7 MBq, the sensitivity of the CZT and LYSO systems are 0.60% and 0.69%, and it increased to 2.43% and 2.55% respectively after recovering MIPEs. The results showed that with a 40 mm crystal thickness, the CZT system could achieve similar sensitivity as the LYSO system. After recovering MIPEs, the sensitivity of both systems improved approximately 3 times, which indicated the importance of MIPE recovery. Due to the poor time resolution, the P-CP sensitivity of the CZT system decreased 25.5% with the increase of activity from 7 MBq to 70 MBq. This was because if more than two interactions were detected within the same time window, all the events within this time window were abandoned. So when the count rate got higher, it was more likely that more than two interactions were detected within the same time window, and lowered the sensitivity.
Fig. 4.

The sensitivity w/o and w/Compton recovery of the CZT and LYSO systems under different source activities. The crystal size of both system is 1 mm.
The P-CP coincidences recovery accuracy of the CZT and LYSO systems based on 1-mm and 2-mm crystal is shown in FIG. 5. The results also showed that given a source activity, the recovery accuracy of the CZT system was about 20% higher than LYSO, and the recovery accuracy was not affected by the source activity. By decreasing the crystal size from 2 mm to 1 mm, the Compton recovery accuracy of the CZT system improved 4.23%, while that of the LYSO system was almost the same. This was because the Compton recovery accuracy was affected by both energy resolution and crystal size. For the CZT system, the energy resolution was high, so decreasing crystal size could improve accuracy. However, in the LYSO system, the poor energy resolution dominated the recovery error, so changing crystal size did not make an obvious influence. Details can be found in our previous study [58].
Fig. 5.

The Compton recovery accuracy of the CZT and LYSO systems based on 1-mm and 2-mm crystal size respectively.
The NEC, true, random, scatter, and the total rate of the CZT and LYSO systems are shown in FIG. 6. The LYSO system showed a high NEC rate at all concentration activities, and the higher the concentration activity, the larger the difference was.
Fig. 6.

The NEC, true, random, scatter, and the total rate of the CZT and LYSO systems. The concentration activity only refers to the neck phantom, while that of the brain and the torso phantoms are kept as the same during the sweeping.
The in-panel and orthogonal-panel spatial resolution of CZT and LYSO systems based on only P-P coincidences and both P-P and P-CP coincidences are shown in Table. I. The results indicated that CZT had a better spatial resolution than LYSO, and incorporating P-CP coincidences for image reconstruction would slightly deteriorate the spatial resolution.
TABLE I.
Point source spatial resolution of the CZT and LYSO system.
| spatial resolution | P-P coincidences | P-P and P-CP coincidences |
|---|---|---|
| CZT in-panel | 0.56 mm | 0.58 mm |
| CZT orthogonal-panel | 0.71 mm | 0.74 mm |
| LYSO in-panel | 0.62 mm | 0.70 mm |
| LYSO orthogonal-panel | 1.31 mm | 1.40 mm |
B. System geometry
The reconstructed images of the cylinder phantom are shown in FIG. 7. The 4, 6, and 8-mm hot spheres were clearly resolvable in all four systems. Due to the limited angular data sampling, the background and hot spheres were elongated in the two-panel and lengthened two-panel systems. CRC and CNR versus hot sphere diameters curves are shown in FIG. 8.
Fig. 7.

Reconstructed images of the cylinder phantom with hot spheres. The top row is the results based on 1-mm crystal size, and the voxel size of the reconstructed image is 0.5×0.5×0.5 mm3. The bottom row is the results based 2-mm crystal size, and the voxel size is 1×1×1 mm3. From left to right, the four systems are two-panel, lengthened two-panel, four-panel, and full-ring, respectively.
Fig. 8.

The contrast recovery coefficient (top) and contrast-to-noise ratio (bottom) versus hot sphere diameter for different system geometries and different crystal size.
The reconstructed images of the XCAT phantom are shown in FIG. 9. The spinal cord, salivary gland, and brain were clearly resolvable in both images. The tumor was also visible in both systems within 6 min, but the tumor in the four-panel system was more resolvable. The tumor CRC were 0.08 and 0.12 for the lengthened two-panel system and the four-panel system, respectively, while the CNR of the tumor were 2.4 and 4.2.
Fig. 9.

Reconstructed images of the XCAT phantom in head and neck region. Top: concentration activity map. Bottom left: the lengthened two-panel system. Bottom right: the four-panel system.
C. Comparison with a whole-body scanner
The transverse slices and the sagittal slice reconstructed images of the dedicated two-panel scanner and the dedicated scanner are shown in FIG. 10. It can be shown that the dedicated system can achieve superior spatial resolution than the whole body system, which indicates the benefits of using such an add-on system.
Fig. 10.

Comparison of the two-panel dedicated system and a whole-body scanner (GE Discovery MI 4-ring PET scanner). (a). the transverse slice of the dedicated scanner; (b). the transverse slice of the whole body scanner. (c). the sagittal slice of the dedicated scanner; (d). the sagittal slice of the whole body scanner.
IV. Discussion
We studied the design considerations for a dedicated HNC PET scanner. Different detecting material including CZT and LYSO, and different system geometries including two-panel, lengthened two-panel, four-panel, and full-ring geometries were investigated.
A. Detecting material
For the detecting material, CZT and LYSO were compared in terms of sensitivity, MIPEs recovery, NEC rate, and spatial resolution. The sensitivity of CZT was 13.04% lower than the LYSO, which suggests that the CZT system and the LYSO system with exactly the same geometry can achieve comparable sensitivity. After recovering MIPEs, the sensitivity of CZT was only 4.71% lower, which indicates that the CZT could recover more MIPEs than LYSO. This is because CZT has a larger Compton-to-photoelectric ratio than the LYSO. Moreover, due to the much better energy resolution, CZT showed an about 20% higher recovery accuracy. The large Compton-to-photoelectric ratio and the high energy resolution make CZT better for the MIPEs recovery.
Though the LYSO system had a larger scatter event rate, it showed a high NEC rate due to its larger true event rate and smaller random event rate. In the data processing, if photons from more than one annihilation were detected in a time window, all singles in this window were rejected. When the total activity was high, the probability to detect more annihilation photons within one window got increased. Since CZT had a wide time window (15 ns), the true coincidence rate of CZT gradually plateaued. The poor time resolution also led to a much higher random rate for the CZT system compared to the LYSO system. The low true rate and high random rate caused the NEC rate for the CZT system to be lower than that of the LYSO system. A lead shield for stopping singles from outside of the FOV is likely necessary for the CZT system.
Table. I shows that the in-panel and orthogonal-panel spatial resolution of the CZT decreased 3.57% and 4.23% after incorporating P-CP coincidences. As a comparison, the LYSO system decreased by 12.90% and 6.87% respectively. The spatial resolution gets worse because incorrectly recovered P-CP coincidences were used for image reconstruction. However, since LYSO has a worse recovery accuracy, the deterioration of spatial resolution is worse than CZT.
Table. I also shows that the CZT system has a better in-panel and orthogonal-panel spatial resolution than LYSO. For a stationary two-panel geometry, DOI affects both the in-panel and orthogonal-panel spatial resolution [18], which is illustrated in FIG. 11. For the sake of simplicity, a 2D case is drawn. The spatial resolution is limited by the closest distinguishable LORs. For two adjacent LORs, the blurring along the orthogonal-panel axis σo is equal to DOI, while the blurring along the in-panel axis σi is equal to DOI × tan(θ), where θ is the angle between the LOR and the orthogonal-panel axis. In a stationary two-panel geometry system, the LOR with a small θ has a large probability to be detected. As a result, σo tends to be larger than σi, which explains why the orthogonal-panel spatial resolution is worse than the in-panel spatial resolution. The result shows the importance of DOI resolution to a two-panel geometry system.
Fig. 11.

Illustration of how DOI affects the in-panel and orthogonal-panel spatial resolution in a stationary two-panel geometry PET system.
B. System geometry
For the system geometry, four CZT systems with two-panel, lengthened two-panel, four-panel, and full-ring geometries were compared in terms of CRC and CNR of the reconstructed image. For all four scanners, both CRC and CNR improved with increasing sphere diameter, and the difference in CRC and CNR among all scanners increased as the diameter of the sphere increased. For all four geometries, the CRC and CNR of 1-mm crystal size were higher than the 2-mm crystal size, which indicated the benefits of using 1-mm crystal.
Compared with the two-panel scanner, the lengthened two-panel scanner had a higher CRC and CNR in 4-, 6- and 8-mm spheres. Compared with the two limited-angle (two-panel, lengthened two-panel) scanners, the two full-geometry (four-panel, full-ring) scanners could achieve higher CRC and CNR in all spheres. However, the improvement of CRC and CNR comes with the need for more detectors. Specifically, the ratio of the number of detectors of the two-panel, lengthened two-panel, four-panel, and full-ring geometries are 10:16:20:23, which means that the full-geometry (four-panel, full-ring) designs achieve higher CRC and CNR in large lesions with a doubled cost, due to the doubled number of detectors.
Moreover, panel-based designs have the flexibility to adjust the panel distance while the same image reconstruction method can still be used, so that the system geometry is able to be compactly adapted to individual patients. It is also very important to consider the patient’s comfort in such a compact design, and two-panel can achieve this goal without blocking the line of sight.
C. Comparison with virtual-pinhole PET
Virtual-pinhole PET is known as using a high-resolution add-on PET scanner to improve the spatial resolution. Depends on different regions of interest, virtual-pinhole PET can have different geometries (full-ring for small animal imaging [59], half-ring for head and neck imaging [60] and breast imaging [61].) The idea of the virtual-pinhole PET and the dedicated PET proposed in this work are similar, both of which use smaller crystals and are placed near the region of interest to improve the spatial resolution locally.
However, the main difference is that the virtual-pinhole PET is inserted into the whole-body PET and it acquires data with the whole body PET simultaneously (a task that is not trivial and requires working with whole-body vendors), thus there are three types of coincidence events: insert-insert, insert-scanner, and scanner-scanner [62]. As a result, the insertion of the virtual-pinhole PET affects the photon detection of the whole-body PET and image reconstruction is more complex because two systems need to be modeled at the same time. As a comparison, the dedicated PET acquires data with whole-body PET separately, so photon detection of the whole-body PET is not affected. However, since we want to use the whole-body PET image as the prior image to reduce the limited artifacts in the dedicated PET image, it is necessary to track patient movement for the dedicated PET application, which brings extra complexity.
D. Limited-angle artifacts
The stationary two-panel PET system is known to have the limited-angle artifacts [63]–[67]. FIG. 7 shows that both the two-panel and lengthened two-panel systems have an elongated background and hot spheres. One method to reduce the limited-angle artifacts is to use a prior image without limited angle artifacts and penalize the dissimilarity between the target image and the prior image during the reconstruction [68], [69]. We plan to take advantage of a whole-body PET scan, which does not have the problem of limited-angle artifacts and the whole-body PET image can be used as the prior image. We are developing the penalized maximum-likelihood image reconstruction algorithm.
V. Conclusions
In this paper, we evaluate different detecting materials and system geometries through the simulation of a dedicated HNC PET scanner. For comparing the detecting material, the sensitivity of the 40-mm thickness CZT system and the 20-mm LYSO system are 0.60% and 0.69%, and it increases to 2.43% and 2.55% respectively after recovering MIPEs. The favorable energy resolution makes the CZT system have an approximately 20% higher MIPE recovery accuracy, but the poor time resolution of CZT results in a two-times lower NEC rate. However, the superior DOI resolution of CZT leads to a better spatial resolution, which is important for HNC imaging. For comparing system geometry, the four-panel and full-ring PET systems achieve higher CRC and CNR than the two-panel and lengthened two-panel PET systems. Nevertheless, the CRC of the 3-mm diameter hot sphere of the two-panel and lengthened two-panel PET systems is comparable to that of the four-panel and full-ring PET systems. The CRC and CNR for the two-panel system can be improved by extending the panel size. Both the two-panel and four-panel PET systems can image the patient’s head and neck region and resolve a 5-mm lesion within 6 min. The disadvantage of the four-panel and full-ring PET systems is that they are in the line of sight of the patient, which compromises the patient’s comfort. To summarize, the CZT system can achieve better spatial resolution and recovery accuracy of MIPEs compared to the LYSO system. However, the poor timing resolution of the CZT system yields a lower sensitivity, especially at high source activity. For system geometry, full-ring and four-panel designs have better CRC and CNR than two-panel design, but they come with a higher cost and compromised patient’s comfort.
Acknowledgment
The authors would also like to thank the Molecular Imaging Instrumentation Laboratory at Stanford University for help with the gpurecon program. This work is supported by NIH R01-EB028091 grant and Cancer Scholars for Translational and Applied Research Program(C*STAR) jointly funded by the Carle Foundation Hospital and the University of Illinois.
Contributor Information
Mohan Li, Department of Nuclear, Plasma, and Radiological Engineering, University of Illinois at Urbana-Champaign, Urbana, IL, 61801 USA..
Brett Yockey, Carle Foundation Hospital, 611 W. Park Street, Urbana, Illinois 61801, USA.
Shiva Abbaszadeh, Department of Nuclear, Plasma, and Radiological Engineering, University of Illinois at Urbana-Champaign, Urbana, IL, 61801 USA..
References
- [1].Siegel RL, Miller KD, and Jemal A, “Cancer statistics, 2017,” CA: A Cancer Journal for Clinicians, vol. 67, pp. 7–30, 2017. [DOI] [PubMed] [Google Scholar]
- [2].Subramaniam R, Truong M, Peller P, Sakai O, and Mercier G, “Fluorodeoxyglucose–positron-emission tomography imaging of head and neck squamous cell cancer,” American Journal of Neuroradiology, vol. 31, no. 4, pp. 598–604, 2010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [3].Nowak B, Di Martino E, Jänicke S, Cremerius U, Adam G, Zimny M, Reinartz P, Büll U et al. , “Diagnostic evaluation of malignant head and neck cancer by F-18-FDG PET compared to CT/MRI,” Nuklearmedizin Archive, vol. 38, no. 8, pp. 312–318, 1999. [PubMed] [Google Scholar]
- [4].Schwartz DL, Ford E, Rajendran J, Yueh B, Coltrera MD, Virgin J, Anzai Y, Haynor D, Lewellyn B, Mattes D et al. , “FDG-PET/CT imaging for preradiotherapy staging of head-and-neck squamous cell carcinoma,” International Journal of Radiation Oncology* Biology* Physics, vol. 61, no. 1, pp. 129–136, 2005. [DOI] [PubMed] [Google Scholar]
- [5].Eschmann S-M, Paulsen F, Reimold M, Dittmann H, Welz S, Reischl G, Machulla H-J, and Bares R, “Prognostic impact of hypoxia imaging with 18F-misonidazole PET in non-small cell lung cancer and head and neck cancer before radiotherapy,” Journal of nuclear medicine, vol. 46, no. 2, pp. 253–260, 2005. [PubMed] [Google Scholar]
- [6].Majdoub M, Hoeben BA, Troost EG, Oyen WJ, Kaanders JH, Le Rest CC, Visser EP, Visvikis D, and Hatt M, “Prognostic Value of Head and Neck Tumor Proliferative Sphericity From 3-Deoxy-3-[18 F] Fluorothymidine Positron Emission Tomography,” IEEE Transactions on Radiation and Plasma Medical Sciences, vol. 2, no. 1, pp. 33–40, 2017. [Google Scholar]
- [7].Adams S, Baum RP, Stuckensen T, Bitter K, and Hör G, “Prospective comparison of 18 F-FDG PET with conventional imaging modalities (CT, MRI, US) in lymph node staging of head and neck cancer,” European Journal of Nuclear Medicine and Molecular Imaging, vol. 25, no. 9, pp. 1255–1260, 1998. [DOI] [PubMed] [Google Scholar]
- [8].Lonsdale MN and Beyer T, “Dual-modality PET/CT instrumentation-today and tomorrow,” European journal of radiology, vol. 73, no. 3, pp. 452–460, 2010. [DOI] [PubMed] [Google Scholar]
- [9].Yao R, Lecomte R, and Crawford ES, “Small-animal PET: what is it, and why do we need it?” Journal of nuclear medicine technology, pp. jnmt–111, 2012. [DOI] [PubMed] [Google Scholar]
- [10].Hoffman EJ, Huang S-C, and Phelps ME, “Quantitation in positron emission computed tomography: 1. Effect of object size.” Journal of computer assisted tomography, vol. 3, no. 3, pp. 299–308, 1979. [DOI] [PubMed] [Google Scholar]
- [11].Hofheinz F, Lougovski A, Zöphel K, Hentschel M, Steffen IG, Apostolova I, Wedel F, Buchert R, Baumann M, Brenner W et al. , “Increased evidence for the prognostic value of primary tumor asphericity in pretherapeutic FDG PET for risk stratification in patients with head and neck cancer,” European journal of nuclear medicine and molecular imaging, vol. 42, no. 3, pp. 429–437, 2015. [DOI] [PubMed] [Google Scholar]
- [12].Yamazaki Y, Saitoh M, Notani K.-i, Tei K, Totsuka Y, Takinami S.-i., Kanegae K, Inubushi M, Tamaki N, and Kitagawa Y, “Assessment of cervical lymph node metastases using FDG-PET in patients with head and neck cancer,” Annals of nuclear medicine, vol. 22, no. 3, pp. 177–184, 2008. [DOI] [PubMed] [Google Scholar]
- [13].Rodrigues RS, Bozza FA, Christian PE, Hoffman JM, Butterfield RI, Christensen CR, Heilbrun M, Wiggins RH, Hunt JP, Bentz BG et al. , “Comparison of whole-body PET/CT, dedicated high-resolution head and neck PET/CT, and contrast-enhanced CT in preoperative staging of clinically M0 squamous cell carcinoma of the head and neck,” Journal of Nuclear Medicine, vol. 50, no. 8, pp. 1205–1213, 2009. [DOI] [PubMed] [Google Scholar]
- [14].Yamamoto Y, Wong TZ, Turkington TG, Hawk TC, and Coleman RE, “Head and neck cancer: dedicated FDG PET/CT protocol for detectionphantom and initial clinical studies,” Radiology, vol. 244, no. 1, pp. 263–272, 2007. [DOI] [PubMed] [Google Scholar]
- [15].Argiris A, Karamouzis MV, Raben D, and Ferris RL, “Head and neck cancer,” The Lancet, vol. 371, no. 9625, pp. 1695–1709, 2008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [16].Carson RE and Kuo PH, “Brain-Dedicated Emission Tomography Systems: A Perspective on Requirements for Clinical Research and Clinical Needs in Brain Imaging,” IEEE Transactions on Radiation and Plasma Medical Sciences, vol. 3, no. 3, pp. 254–261, 2019. [Google Scholar]
- [17].Hutton B, Yamaya T, and Furenlid L, “Dedicated Molecular Imaging Systems for Human Neurological Studies,” IEEE Transactions on Radiation and Plasma Medical Sciences, vol. 3, no. 3, pp. 252–253, 2019. [Google Scholar]
- [18].Peng H and Levin CS, “Design study of a high-resolution breast-dedicated PET system built from cadmium zinc telluride detectors,” Physics in medicine and biology, vol. 55, no. 9, p. 2761, 2010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [19].Abbaszadeh S, Gu Y, Reynolds PD, and Levin CS, “Characterization of a sub-assembly of 3D position sensitive cadmium zinc telluride detectors and electronics from a sub-millimeter resolution pet system,” Physics in Medicine & Biology, vol. 61, no. 18, p. 6733, 2016. [DOI] [PubMed] [Google Scholar]
- [20].Drezet A, Monnet O, Mathy F, Montemont G, and Verger L, “CdZnTe detectors for small field of view positron emission tomographic imaging,” Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, vol. 571, no. 1–2, pp. 465–470, 2007. [Google Scholar]
- [21].Vaska P, Dragone A, Lee W, Kim D-H, Pratte J-F, Cui Y-G, Fried J, Krishnamoorthy S, Bolotnikov A, Park S-J et al. , “A prototype CZT-based PET scanner for high resolution mouse brain imaging,” in Nuclear Science Symposium Conference Record, 2007. NSS’07. IEEE, vol. 5 IEEE, 2007, pp. 3816–3819. [Google Scholar]
- [22].Mitchell GS, Sinha S, Stickel JR, Bowen SL, Cirignano LJ, Dokhale P, Kim H, Shah KS, and Cherry SR, “CdTe strip detector characterization for high resolution small animal PET,” IEEE Transactions on Nuclear Science, vol. 55, no. 3, pp. 870–876, 2008. [Google Scholar]
- [23].Morimoto Y, Ueno Y, Takeuchi W, Kojima S, Matsuzaki K, Ishitsu T, Umegaki K, Kiyanagi Y, Kubo N, Katoh C et al. , “Development of a 3D brain PET scanner using CdTe semiconductor detectors and its first clinical application,” IEEE Transactions on Nuclear Science, vol. 58, no. 5, pp. 2181–2189, 2011. [Google Scholar]
- [24].González AJ, Sánchez F, and Benlloch JM, “Organ-dedicated molecular imaging systems,” IEEE Transactions on Radiation and Plasma Medical Sciences, vol. 2, no. 5, pp. 388–403, 2018. [Google Scholar]
- [25].Bauer CE, Brefczynski-Lewis J, Marano G, Mandich M-B, Stolin A, Martone P, Lewis JW, Jaliparthi G, Raylman RR, and Majewski S, “Concept of an upright wearable positron emission tomography imager in humans,” Brain and behavior, vol. 6, no. 9, 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [26].Yamamoto S, Honda M, Oohashi T, Shimizu K, and Senda M, “Development of a brain PET system, PET-Hat: a wearable PET system for brain research,” IEEE Transactions on Nuclear Science, vol. 58, no. 3, pp. 668–673, 2011. [Google Scholar]
- [27].Tashima H, Yoshida E, Nishikido F, Wakizaka H, Nitta M, Ahmed AM, Mohammadi A, Tazawa S, Kimura Y, Suhara T et al. , “Development of the helmet-chin PET prototype,” in 2015 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC). IEEE, 2015, pp. 1–3. [Google Scholar]
- [28].De Jong HW, Van Velden FH, Kloet RW, Buijs FL, Boellaard R, and Lammertsma AA, “Performance evaluation of the ECAT HRRT: an LSO-LYSO double layer high resolution, high sensitivity scanner,” Physics in Medicine & Biology, vol. 52, no. 5, p. 1505, 2007. [DOI] [PubMed] [Google Scholar]
- [29].MacDonald L, Edwards J, Lewellen T, Haseley D, Rogers J, and Kinahan P, “Clinical imaging characteristics of the positron emission mammography camera: PEM Flex Solo II,” Journal of Nuclear Medicine, vol. 50, no. 10, pp. 1666–1675, 2009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [30].Ferrero A, Peng Q, Burkett GW Jr, Sumanasena B, Moses WW, and Badawi RD, “Preliminary performance characterization of DbPET2. 1, a PET scanner dedicated to the imaging of the breast and extremities,” Biomedical Physics & Engineering Express, vol. 1, no. 1, p. 015202, 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [31].Miyake KK, Matsumoto K, Inoue M, Nakamoto Y, Kanao S, Oishi T, Kawase S, Kitamura K, Yamakawa Y, Akazawa A et al. , “Performance evaluation of a new dedicated breast PET scanner using NEMA NU4–2008 standards,” Journal of Nuclear Medicine, vol. 55, no. 7, pp. 1198–1203, 2014. [DOI] [PubMed] [Google Scholar]
- [32].Moliner L, Gonzalez A, Soriano A, Sánchez F, Correcher C, Orero A, Carles M, Vidal L, Barbera J, Caballero L et al. , “Design and evaluation of the MAMMI dedicated breast PET,” Medical physics, vol. 39, no. 9, pp. 5393–5404, 2012. [DOI] [PubMed] [Google Scholar]
- [33].Jan S, Santin G, Strul D, Staelens S, Assie K, Autret D, Avner S, Barbier R, Bardies M, Bloomfield P et al. , “GATE: a simulation toolkit for PET and SPECT,” Physics in medicine and biology, vol. 49, no. 19, p. 4543, 2004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [34].Lee J-H, Hwang S, Istook CL et al. , “Analysis of human head shapes in the united states,” International Journal of Human Ecology, vol. 7, no. 1, pp. 77–83, 2006. [Google Scholar]
- [35].Sportelli G, Ahmad S, Belcari N, Bisogni MG, Camarlinghi N, Di Pasquale A, Dussoni S, Fleury J, Morrocchi M, Zaccaro E et al. , “The TRIMAGE PET Data Acquisition System: Initial Results,” IEEE Transactions on Radiation and Plasma Medical Sciences, vol. 1, no. 2, pp. 168–177, 2017. [Google Scholar]
- [36].Majewski S, Proffitt J, Brefczynski-Lewis J, Stolin A, Weisenberger A, Xi W, and Wojcik R, “HelmetPET: A silicon photomultiplier based wearable brain imager,” in Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE. IEEE, 2011, pp. 4030–4034. [Google Scholar]
- [37].Li M and Abbaszadeh S, “Depth-of-interaction study of a dual-readout detector based on TOFPET2 application-specific integrated circuit,” Physics in Medicine & Biology, 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [38].Di Francesco A, Bugalho R, Oliveira L, Pacher L, Rivetti A, Rolo M, Silva J, Silva R, and Varela J, “TOFPET2: a high-performance ASIC for time and amplitude measurements of SiPM signals in time-of-flight applications,” Journal of Instrumentation, vol. 11, no. 03, p. C03042, 2016. [Google Scholar]
- [39].Cajipe VB, Clajus M, Hayakawa S, Matteson JL, Skelton RT, Tumer TO, and Volkovskii A, “Performance of the RENA-3 IC with position-sensitive CZT and CdTe detectors,” in 2008 IEEE Nuclear Science Symposium Conference Record. IEEE, 2008, pp. 300–307. [Google Scholar]
- [40].Abbaszadeh S, Chinn G, and Levin CS, “Positioning true coincidences that undergo inter-and intra-crystal scatter for a sub-mm resolution cadmium zinc telluride-based PET system,” Physics in Medicine & Biology, vol. 63, no. 2, p. 025012, 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [41].Gillam JE, Solevi P, Oliver JF, Casella C, Heller M, Joram C, and Rafecas M, “Sensitivity recovery for the AX-PET prototype using inter-crystal scattering events,” Physics in medicine and biology, vol. 59, no. 15, p. 4065, 2014. [DOI] [PubMed] [Google Scholar]
- [42].Wagadarikar AA, Ivan A, Dolinsky S, and McDaniel DL, “Sensitivity improvement of time-of-flight (TOF)-PET detector through recovery of Compton scattered annihilation photons,” in Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2012 IEEE. IEEE, 2012, pp. 3178–3183. [Google Scholar]
- [43].Yoshida E, Tashima H, and Yamaya T, “Sensitivity booster for DOI-PET scanner by utilizing Compton scattering events between detector blocks,” Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, vol. 763, pp. 502–509, 2014. [Google Scholar]
- [44].Yoshida E, Kobayashi A, Yamaya T, Watanabe M, Nishikido F, Kitamura K, Hasegawa T, Fukushi M, and Murayama H, “The jPETD4: performance evaluation of four-layer DOI-PET scanner using the NEMA NU2–2001 standard,” in 2006 IEEE Nuclear Science Symposium Conference Record, vol. 4 IEEE, 2006, pp. 2532–2536. [Google Scholar]
- [45].Wienhard K, Schmand M, Casey M, Baker K, Bao J, Eriksson L, Jones W, Knoess C, Lenox M, Lercher M et al. , “The ECAT HRRT: performance and first clinical application of the new high resolution research tomograph,” IEEE Transactions on Nuclear Science, vol. 49, no. 1, pp. 104–110, 2002. [Google Scholar]
- [46].Gong K, Majewski S, Kinahan PE, Harrison RL, Elston BF, Manjeshwar R, Dolinsky S, Stolin AV, Brefczynski-Lewis JA, and Qi J, “Designing a compact high performance brain PET scannersimulation study,” Physics in medicine and biology, vol. 61, no. 10, p. 3681, 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [47].Strother S, Casey M, and Hoffman E, “Measuring PET scanner sensitivity: relating countrates to image signal-to-noise ratios using noise equivalents counts,” IEEE Transactions on Nuclear Science, vol. 37, no. 2, pp. 783–788, 1990. [Google Scholar]
- [48].Engel H, Steinert H, Buck A, Berthold T, Huch Böni R, and Von Schulthess GK, “Whole-body PET: physiological and artifactual fluorodeoxyglucose accumulations,” Journal of Nuclear Medicine, vol. 37, no. 3, pp. 441–445, 1996. [PubMed] [Google Scholar]
- [49].Shepp LA and Vardi Y, “Maximum likelihood reconstruction for emission tomography,” IEEE transactions on medical imaging, vol. 1, no. 2, pp. 113–122, 1982. [DOI] [PubMed] [Google Scholar]
- [50].Cui J, Pratx G, Meng B, and Levin CS, “Distributed MLEM: An iterative tomographic image reconstruction algorithm for distributed memory architectures,” IEEE transactions on medical imaging, vol. 32, no. 5, pp. 957–967, 2013. [DOI] [PubMed] [Google Scholar]
- [51].Vandenberghe S, Mikhaylova E, DHoe E, Mollet P, and Karp JS, “Recent developments in time-of-flight PET,” EJNMMI physics, vol. 3, no. 1, p. 3, 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [52].Surti S and Karp JS, “Advances in time-of-flight PET,” Physica medica, vol. 32, no. 1, pp. 12–22, 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [53].——, “Design considerations for a limited angle, dedicated breast, TOF PET scanner,” Physics in Medicine & Biology, vol. 53, no. 11, p. 2911, 2008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [54].Segars WP, Sturgeon G, Mendonca S, Grimes J, and Tsui BMW, “4D XCAT phantom for multimodality imaging research,” Medical physics, vol. 37, no. 9, pp. 4902–4915, 2010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [55].Nakamoto Y, Tatsumi M, Hammoud D, Cohade C, Osman MM, and Wahl RL, “Normal FDG distribution patterns in the head and neck: PET/CT evaluation,” Radiology, vol. 234, no. 3, pp. 879–885, 2005. [DOI] [PubMed] [Google Scholar]
- [56].Higgins KA, Hoang JK, Roach MC, Chino J, Yoo DS, Turkington TG, and Brizel DM, “Analysis of pretreatment FDG-PET SUV parameters in head-and-neck cancer: tumor SUVmean has superior prognostic value,” International Journal of Radiation Oncology*Biology*Physics, vol. 82, no. 2, pp. 548–533, 2012. [DOI] [PubMed] [Google Scholar]
- [57].Engel H, Steinert H, Buck A, Berthold T, Huch Böni RA, and von Schulthess GK, “Whole-body PET: physiological and artifactual fluorodeoxyglucose accumulations,” Journal of Nuclear Medicine, vol. 37, no. 3, pp. 441–445, 1996. [PubMed] [Google Scholar]
- [58].Yang S, Li M, Reed M, Hugg J, Chen H, and Abbaszadeh S, “Effect of CZT system characteristics on Compton scatter event recovery,” IEEE Transactions on Radiation and Plasma Medical Sciences, 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [59].Wu H, Pal D, Song TY, O’Sullivan JA, and Tai Y-C, “Micro insert: A prototype full-ring PET device for improving the image resolution of a small-animal PET scanner,” Journal of nuclear medicine, vol. 49, no. 10, pp. 1668–1676, 2008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [60].Wu H, Song TY, Pal D, Keesing DB, Komarov S, Sullivan JA, and Tai Y-C, “A high resolution PET insert system for clinical PET/CT scanners,” in 2008 IEEE Nuclear Science Symposium Conference Record. IEEE, 2008, pp. 5442–5444. [Google Scholar]
- [61].Mathews AJ, Komarov S, Wu H, O’Sullivan JA, and Tai Y-C, “Improving PET imaging for breast cancer using virtual pinhole PET half-ring insert,” Physics in Medicine & Biology, vol. 58, no. 18, p. 6407, 2013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [62].Keesing DB, Mathews A, Komarov S, Wu H, Song TY, OSullivan JA, and Tai Y-C, “Image reconstruction and system modeling techniques for virtual-pinhole PET insert systems,” Physics in Medicine & Biology, vol. 57, no. 9, p. 2517, 2012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [63].Lee E, Werner ME, Karp JS, and Surti S, “Design optimization of a time-of-flight, breast PET scanner,” IEEE transactions on nuclear science, vol. 60, no. 3, pp. 1645–1652, 2013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [64].Doshi NK, Silverman RW, Shao Y, and Cherry SR, “maxPET, a dedicated mammary and axillary region PET imaging system for breast cancer,” IEEE Transactions on Nuclear Science, vol. 48, no. 3, pp. 811–815, 2001. [Google Scholar]
- [65].Raylman RR, Majewski S, Smith MF, Proffitt J, Hammond W, Srinivasan A, McKisson J, Popov V, Weisenberger A, Judy CO et al. , “The positron emission mammography/tomography breast imaging and biopsy system (PEM/PET): design, construction and phantom-based measurements,” Physics in Medicine & Biology, vol. 53, no. 3, p. 637, 2008. [DOI] [PubMed] [Google Scholar]
- [66].Smith MF, Raylman RR, Majewski S, and Weisenberger AG, “Positron emission mammography with tomographic acquisition using dual planar detectors: initial evaluations,” Physics in Medicine & Biology, vol. 49, no. 11, p. 2437, 2004. [DOI] [PubMed] [Google Scholar]
- [67].Abreu MC, Aguiar JD, Almeida FG, Almeida P, Bento P, Carriço B, Ferreira M, Ferreira NC, Gonçalves F, Leong C et al. , “Design and evaluation of the Clear-PEM scanner for positron emission mammography,” IEEE Transactions on Nuclear Science, vol. 53, no. 1, pp. 71–77, 2006. [Google Scholar]
- [68].Chen G-H, Tang J, and Leng S, “Prior image constrained compressed sensing (PICCS): a method to accurately reconstruct dynamic CT images from highly undersampled projection data sets,” Medical physics, vol. 35, no. 2, pp. 660–663, 2008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [69].Chen G-H and Li Y, “Synchronized multiartifact reduction with tomographic reconstruction (SMART-RECON): A statistical model based iterative image reconstruction method to eliminate limited-view artifacts and to mitigate the temporal-average artifacts in time-resolved ct,” Medical physics, vol. 42, no. 8, pp. 4698–4707, 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]