Table I. Single injection DIA acquisition settings for quantitative proteomics with Thermo instruments.
QE | QE+ | QE-HF | QE-HFX | Fusion | Lumos | |
---|---|---|---|---|---|---|
Est. Scan Rate | 10 Hz | 10 Hz | 10 Hz | 20 Hz | 10 Hz | 20 Hz |
MS1 Scans | ||||||
Range | 385–1015 m/z | 385–1015 m/z | 385–1015 m/z | 390–1010 m/z | 385–1015 m/z | 390–1010 m/z |
Resolution | 35000 | 35000 | 60000 | 60000 | 60000 (Orbi) | 60000 (Orbi) |
Max IIT (ms) | 60 | 60 | 60 | 60 | 60 | 60 |
AGC Target | 1E+6 | 1E+6 | 1E+6 | 1E+6 | 4E+5 | 4E+5 |
Data Type | Centroid | Centroid | Centroid | Centroid | Centroid | Centroid |
MS2 Scans | ||||||
Windowing Scheme | 24 m/z staggered | 24 m/z staggered | 24 m/z staggered | 8 m/z staggered | 24 m/z staggered | 8 m/z staggered |
Resolution | 17500 | 17500 | 30000 | 15000 | 30000 (Orbi) | 15000 (Orbi) |
Max IIT (ms) | 60 | 60 | 60 | 20 | 60 | 20 |
AGC Target | 1e6 | 1e6 | 1e6 | 1e6 | 4e5 | 4e5 |
Loop Count | 25 | 25 | 25 | 75 | 25 | 75 |
Default Charge | 3 | 3 | 3 | 3 | 3 | 3 |
(N)CE | 27 | 27 | 27 | 27 | 33 (HCD) | 33 (HCD) |
Data Type | Centroid | Centroid | Centroid | Centroid | Centroid | Centroid |