Table II. 6× GPF-DIA acquisition settings for DIA-only library generation with Thermo instruments.
QE | QE+ | QE-HF | QE-HFX | Fusion | Lumos | |
---|---|---|---|---|---|---|
Est. Scan Rate | 10 Hz | 10 Hz | 10 Hz | 10 Hz | 10 Hz | 10 Hz |
MS1 Scans | ||||||
Range | 395–505 m/z, 495–605 m/z, etc… | 395–505 m/z, 495–605 m/z, etc… | 395–505 m/z, 495–605 m/z, etc… | 395–505 m/z, 495–605 m/z, etc… | 395–505 m/z, 495–605 m/z, etc… | 395–505 m/z, 495–605 m/z, etc… |
Resolution | 35000 | 35000 | 60000 | 60000 | 60000 (Orbi) | 60000 (Orbi) |
Max IIT (ms) | 60 | 60 | 60 | 60 | 60 | 60 |
AGC Target | 1E+6 | 1E+6 | 1E+6 | 1E+6 | 4E+5 | 4E+5 |
Data Type | Centroid | Centroid | Centroid | Centroid | Centroid | Centroid |
MS2 Scans | ||||||
Windowing Scheme | 6x GPF-DIA w/4 m/z staggered | 6x GPF-DIA w/4 m/z staggered | 6x GPF-DIA w/4 m/z staggered | 6x GPF-DIA w/4 m/z staggered | 6x GPF-DIA w/4 m/z staggered | 6x GPF-DIA w/4 m/z staggered |
Resolution | 17500 | 17500 | 30000 | 30000 | 30000 (Orbi) | 30000 (Orbi) |
Max IIT (ms) | 60 | 60 | 60 | 60 | 60 | 60 |
AGC Target | 1e6 | 1e6 | 1e6 | 1e6 | 4e5 | 4e5 |
Loop Count | 25 | 25 | 25 | 25 | 25 | 25 |
Default Charge | 3 | 3 | 3 | 3 | 3 | 3 |
(N)CE | 27 | 27 | 27 | 27 | 33 (HCD) | 33 (HCD) |
Data Type | Centroid | Centroid | Centroid | Centroid | Centroid | Centroid |