Skip to main content
. 2020 Jul 7;4:38. doi: 10.1186/s41747-020-00165-2

Table 3.

Estimation of the m values for different tube voltage values

kVp m (cm-1)
26 0.65 ± 0.03
28 0.60 ± 0.02
30 0.55 ± 0.02
32 0.52 ± 0.02
34 0.48 ± 0.02
36 0.46 ± 0.01
38 0.44 ± 0.01
40 0.42 ± 0.01
44 0.39 ± 0.01
48 0.36 ± 0.01

Data for m values are presented as mean ± standard deviation. The m values were obtained by fitting experimental data through Eq. (4), as shown in Fig. 1. The fitting parameters of Eq. (4) are presented in Table 2