Table 4.

Comparison between 2ABD computed from Eq. (1) and the average glandular dose (AGD) calculated by the Dance approach for 100% glandularity

Mammographic device Phantom thickness (cm) Tube load (mAs) Tube voltage (kVp) 2ABD (mGy) AGD 100% glandularity (mGy) Relative difference (%)
Hologic-A 3.0 37.5 28 0.94 ± 0.19 1.00 ± 0.20 -6.0
Hologic-A 3.5 40 29 1.05 ± 0.20 1.07 ± 0.21 -1.9
Hologic-A 4.0 50 29 1.21 ± 0.23 1.21 ± 0.24 0.0
Hologic-A 4.5 50 30 1.26 ± 0.23 1.27 ± 0.25 -0.8
Hologic-A 5.0 55 31 1.45 ± 0.26 1.46 ± 0.29 -0.7
Hologic-A 5.5 60 32 1.66 ± 0.30 1.67 ± 0.33 -0.6
Hologic-A 6.0 65 33 1.87 ± 0.33 1.89 ± 0.38 -1.1
Hologic-A 6.5 70 34 2.12 ± 0.37 2.14 ± 0.43 -0.9
Hologic-A 7.0 80 35 2.51 ± 0.44 2.57 ± 0.51 -2.3
Hologic-B 3.5 30 29 1.07 ± 0.20 1.14 ± 0.23 -6.1
Hologic-B 4.5 35 30 1.20 ± 0.22 1.22 ± 0.24 -1.6
Hologic-B 5.5 42.5 32 1.61 ± 0.29 1.64 ± 0.33 -1.8
Hologic-B 6.5 50 34 2.06 ± 0.37 2.09 ± 0.42 -1.4
Fuji-C 3.5 28 29 1.19 ± 0.24 1.21 ± 0.24 -1.7
Fuji-C 4.5 32 32 1.63 ± 0.30 1.56 ± 0.31 4.5
Fuji-C 5.5 42 33 2.06 ± 0.37 2.04 ± 0.41 1.0
Fuji-C 6.5 50 35 2.66 ± 0.47 2.65 ± 0.53 0.4
Fuji-C 4.5* 45 32 2.30 ± 0.43 2.15 ± 0.43 7.0
Fuji-C 6.5* 63 35 3.35 ± 0.59 3.27 ± 0.65 2.4
Fuji-D 3.5 28 30 1.30 ± 0.26 1.38 ± 0.28 -5.8
Fuji-D 4.5 32 32 1.59 ± 0.30 1.60 ± 0.32 -0.6
Fuji-D 5.5 50 33 2.40 ± 0.43 2.43 ± 0.49 -1.2
Fuji-D 6.5 56 35 2.91 ± 0.50 2.87 ± 0.57 1.4
Fuji-D 3.5* 32 30 1.49 ± 0.28 1.55 ± 0.31 -3.9
Fuji-D 4.5* 45 32 2.24 ± 0.43 2.20 ± 0.44 1.8
Fuji-D 5.5* 63 34 3.35 ± 0.61 3.22 ± 0.64 4.0
Fuji-D 6.5* 71 35 3.69 ± 0.65 3.56 ± 0.71 3.7

Data for 2ABD and AGD 100% glandularity are presented as mean ± standard deviation. The comparison was carried out by considering all mammographic devices. The exposure parameters were chosen as close as possible to the corresponding automatic exposure control settings for a given phantom thickness. For the AGD calculation by the Dance method, the measured air kerma was employed. For the Amulet Innovality devices both the standard acquisition mode and high-resolution acquisition mode (*) were employed