. 2017 Dec 27;122:1–19. doi: 10.6028/jres.122.044

Table 4.

LS counting results for the first trial as obtained in the two counters. The average quench-indicating parameter H# (measured in counter “B”) maps with the composition variations shown in Table 5. Values for the massic counting rate RU/m are grand mean averages for the three sources for each series and across the n = 4 (for “B”) or n = 3 (for “H”) replicate measurements for each source. The relative measurement precision estimator S is the combination in quadrature of the typical within-source standard deviation of the mean for the three sources and the between-source standard deviation of the three mean values.

Series Average H# “B” counter “H” counter
RU/m (s−1 g−1) S (%) RU/m (s−1 g−1) S (%)
A 104.2 905.5 0.17 906.9 0.56
B 120.0 904.5 0.10 905.0 0.14
C 122.9 903.9 0.11 905.0 0.62