Table 1.

Results of least squares fitting of Eq. (3) to LO data for two stock solutions of different microsphere types, on two different LO instruments of the same type. Stated uncertainties are type A standard uncertainty of the mean.

Microsphere lota S S URB URB
Instrument A A A B
Run date 28 AUG 2014 12 MAY 2016 16 NOV 2016 17 NOV 2016
Lower number concentration limit, mL−1 2,000 10,000 5,000 5,000
Upper number concentration limit, mL−1 10,000 46,000 36,000 36,000
αc, mL (1.79 ± 0.38)×10−6 (1.76 ± 0.49)×10−6 (1.81 ± 0.07)×10−6 (1.18 ± 0.08)×10−6
αt −0.0016 ± 0.0010 −0.0026 ± 0.0007 −0.0013 ± 0.0006 −0.00161 ± 0.0007
N, mL−1 95,377 ± 382 100,583 ± 230 96,545 ± 252 95,892 ± 281
Reduced chi-square, χv2 1.12 0.52 1.14 1.45
DF 78 27 45 45
a

S = sulfate functionalized PSL microspheres, URB = fluorescently dyed “ultrarainbow microspheres.”