Skip to main content
. 2020 Jun 29;2020:7457902. doi: 10.1155/2020/7457902

Table 3.

Visual and refractive outcomes, intraocular pressure (IOP), endothelial cell density (ECD), and vault of the different studies considered (mean, standard deviation, and range) NR: not reported.

Author UCVA (logMAR) CDVA (logMAR) Safety index Efficacy index ±0.50D (% eyes) ±1.00D (% eyes) SE (D) IOP (mmHg) (eyes >21) ECD (cells/m2) (% loss) Vault (μm)
Lisa et al. [5] 0.028 ± 0.055 0.003 ± 0.013 1.04 1.00 93.9 100 −0.14 ± 0.26 12.4 ± 1.4 (0 > 20) 2650 ± 438 (1.7) 405.5 ± 184.7 (100 to 980)
Karandikar et al. [6] NR NR 1.15 1.6 57.12 98.12 −0.19 ± 1.18 19.1 ± 1.3 NR (7.1) 628.2 ± 300.1 (NR)
Ganesh et al. [7] −0.022 ± 0.021 −0.071 ± 0.079 1.24 1.12 90 100 −0.164 ± 0.20 NR 2808 ± 315 (9.0) NR
Pjano et al. [8] 0.76 ± 0.16 (decimal) 0.79 ± 0.14 (decimal) 1.25 1.2 NR NR −0.21 ± 0.27 14.96 ± 1.7 2512 ± 127 (5.5) NR
Kamiya et al. [9]
 Low-to-moderate myopia −0.17 ± 0.14 −0.21 ± 0.10 NR NR 93 98 −0.12 13.1 (0) NR (0.1) NR
 High myopia −0.16 ± 0.09 −0.21 ± 0.08 NR NR 94 99 0.02 13.6 (0) NR (0.1) NR
García-de la Rosa et al. [10] 0.12 ± 0.12 0.05 ± 0.08 NR NR 85 86 −0.06 ± 0.77 NR NR 449 ± 180 (NR)
Niu et al. [11] 0.89 ± 0.30 (decimal) 1.00 ± 0.27 (decimal) 1.33 1.14 69 92 −0.67 ± 1.29 15.15 ± 2.57 (0) 2963 ± 396 (8.38) 380 ± 152 (90 to 700)
Sachdev et al. [12] 0 ± NR 0 ± NR NR NR 94.09 96.09 −0.12 ± NR NR NR NR
Current study 0.88 ± 0.16 (decimal) 0.96 ± 0.09 (decimal) 1.09 1.00 90.6 93.9 −0.09 ± 0.47 15.27 ± 3.03 (0) 2516 ± 234 (2.04) 320 ± 136 (96 to 603)

UCVA: uncorrected distance visual acuity, CDVA: corrected distance visual acuity, SE: spherical equivalent, and NR: not reported.