Figure 5.
Comparison of the theoretical real-time urea clearance (see Equation (15)) to two multi-pass benchtop experiments with the sheet membrane device. Agreement was observed to match theory, thus showing that our model for is accurate (see Equation (18)). Amp = 2.88 × 10–4 m2; Asp =1.4 × 10–6 m2. (Experiment 1: R2 = 0.98; Experiment 2: R2 = 0.96.)