Table 2.

Summary of the measurements of mechanical properties.

Sample Tested HV100 Re [MPa] KIC [MPa•m1/2]
Tibial Component, New
Sample 1


Sample 2
365
360
359
360
362
358
Mean
360
Standard Deviation
2.50
122
120
119
120
120
119
Mean
120
Standard Deviation
1.09
108
104
106
105
108
105
Mean
106
Standard Deviation
1.67
Tibial Component Used for 6 Years
Sample 1


Sample 2


Sample 3
376
378
380
384
380
378
377
388
382
Mean
380
Standard Deviation
3.81
125
126
127
128
127
126
126
129
127
Mean
127
Standard Deviation
1.20
97
101
102
100
101
102
99
97
101
Mean
100
Standard Deviation
1.94
Tibial Component Used for 12 Years
Sample 1 432
428
431
Mean
430
Standard Deviation
2.52
144
142
14
Mean
143
Standard Deviation
1
77
77
77
Mean
77
Standard Deviation
0
Unused Tibial Component
Sample 1


Sample 2
397
400
394
407
406
396
Mean
400
Standard Deviation
5.40
132
133
131
136
136
131
Mean
133
Standard Deviation
2.25
130
129
122
125
127
123
Mean
126
Standard Deviation
3.22
Femoral Component Used for 6 Years
Sample 1


Sample 2


Sample 3
411
406
405
411
416
415
407
411
411
Mean
410
Standard Deviation
3.77
137
135
135
137
139
138
135
137
137
Mean
137
Standard Deviation
1.41
120
128
131
124
122
121
123
118
121
Mean
123
Standard Deviation
4.07
Femoral Component Used for 12 Years
Sample 1


437
444
438
Mean
440
Standard Deviation
3.78
145
148
146
Mean
147
Standard Deviation
1.15
100
96
98
Mean
98
Standard Deviation
2