Comparison of various filters available for use in COVID-19 patients requiring KRT.

Characteristics Filters
CytoSorb (Cytosorbents) oXiris (Baxter)
Membrane composition Polystyrene divinylbenzene co-polymer microporous beads (coated with polyvinylpyrrolidone) AN69 copolymer covered with polyethyleneimine and unfractionated heparin
Sterilization type Gamma irradiation Ethylene oxide
Capability of adsorption Cytokines Endotoxin and cytokines
Adsorption mode Hydrophobic interactions Ionic interactions-cytokines due to sulfonate groups.-endotoxins due to high PEI concentration on inner part of membrane.
Heparin-covered inner surface No Yes

Adapted from Karkar and Ronco (29).