Table 2.
Percentage of 2D Retinal Nerve Fiber Layer Scans with Artifacts
| Glaucoma (n = 106) | Normal (n = 95) | ||
|---|---|---|---|
| Artifact Type | Percentage of B-scans (%)a | Percentage of B-scans (%)a | P |
| Posterior RNFL misidentification | 46.2 | 5.3 | <0.001*** |
| Decentration | 10.4 | 8.4 | 0.64 |
| Anterior RNFL misidentification | 9.4 | 3.2 | 0.07 |
| PVD/ERM | 4.7 | 2.1 | 0.32 |
| Degraded image | 2.8 | 0 | 0.10 |
| Peripapillary atrophy | 1.9 | 0 | 0.18 |
| Complete absence of segmentation | 0 | 0 | N/A |
| Out of register | 0 | 0 | N/A |
| Cut edge | 0 | 0 | N/A |
| Mirroring | 0 | 0 | N/A |
| Missing parts | 0 | 0 | N/A |
| Motion artifact | 0 | 0 | N/A |
| Any artifact (total percentage)b | 58.5 | 14.7 | <0.001*** |
Proportion of 2D retinal nerve fiber layer scans with the specified artifact type divided by 106 for the glaucoma group and 95 for the normal group.
Total percentage represents the total percentage of 2D retinal nerve fiber layer scans with at least one artifact type. Scans with multiple artifacts were only counted once.