. 2020 Jun 19;10(6):1196. doi: 10.3390/nano10061196

Table 3.

Precision data for the determination of GEF in water and plasma samples by the proposed method.

Intra-Day Precision Inter-Day Precision
Conc. Taken (ng/mL) Conc. Found (ng/mL) % Found Conc. Taken (ng/mL) Conc. Found (ng/mL) % Found
Water samples 50.00 50.62 101.24 50.00 50.62 101.24
150.00 151.99 101.33 150.00 152.63 101.75
300.00 299.97 99.99 300.00 299.76 99.92
X¯ ± SD 100.85 ± 0.75 100.95 ± 0.94
%RSD 0.75 0.94
%Error 0.43 0.54
Plasma samples 150.00 147.76 98.51 150.00 149.86 99.91
450.00 454.54 101.01 450.00 452.43 100.54
600.00 596.94 99.49 600.00 584.16 97.36
X¯ ± SD 99.67 ± 1.26 99.27 ± 1.70
%RSD 1.26 1.70
%Error 0.73 0.98

N.B. Each result is the average of three separate determinations.