Table 1.
Real-time ventilation feedback device validation data.
| Set Volume of Mechanical Ventilation, mL | Analyser | RTVFD | p-Value |
|---|---|---|---|
| 150 | 149.4 ± 1.0 | 144 ± 9.8 | 0.008 |
| 250 | 253.9 ± 0.3 | 244.3 ± 9.7 | <0.001 |
| 350 | 348.2 ± 0.6 | 348.0 ± 10.9 | 0.903 |
| 450 | 455.8 ± 1.3 | 450.1 ± 9.1 | 0.005 |
| 550 | 551.2 ± 2.2 | 545.2 ± 13.7 | 0.040 |
RTVFD, real-time ventilation feedback device. Values are mean ± SD.