Table 3:
Repeatability coefficient (CR) and variance (σ2) of ADCexp values calculated for different acquisition schemes and acquisition times, in VOIPCC and VOICR. Power analysis can be performed using the σ2 values reported in the table and Equation 4.
VOI | Metabolite | 12 minute acquisition (scheme b[0,1,2]) | 7 minute acquisition (scheme b[0,2]) | 4 minute acquisition (scheme b[0,2]) | |||
---|---|---|---|---|---|---|---|
CR (%mean) (μm2/ms) | σ2 (x10−4) (μm4/ms2) | CR (%mean) (μm2/ms) | σ2 (x10−4) (μm4/ms2) | CR (%mean) (μm2/ms) | σ2 (x10−4) (μm4/ms2) | ||
VOIPCC | tNAA | 0.015 (12%) | 0.3 | 0.026 (20%) | 0.9 | 0.016 (13%) | 0.3 |
tCr | 0.023 (20%) | 0.7 | 0.032 (26%) | 1.3 | 0.030 (25%) | 1.1 | |
tCho | 0.022 (20%) | 0.6 | 0.040 (37%) | 2.1 | 0.025 (23%) | 0.8 | |
mIns | 0.040 (37%) | 2.1 | 0.066 (59%) | 5.7 | 0.050 (46%) | 3.3 | |
VOICR | tNAA | 0.042 (27%) | 2.3 | 0.042 (26%) | 2.3 | 0.045 (27%) | 2.6 |
tCr | 0.040 (24%) | 2.1 | 0.043 (25%) | 2.4 | 0.052 (29%) | 3.5 | |
tCho | 0.044 (33%) | 2.6 | 0.046 (33%) | 2.7 | 0.050 (34%) | 3.2 | |
mIns | 0.104 (75%) | 14.1 | 0.085 (62%) | 9.4 | 0.101 (67%) | 13.4 |