Skip to main content
. 2020 Mar 26;43(7):zsaa045. doi: 10.1093/sleep/zsaa045

Table 3.

Comparison of Night-Level Metrics Derived From Device Data to Those Derived From RPSGT Staging

Metric Device Mean error MAE RMSE R 2
Sleep efficiency (%) ActiGraph Link (Cole–Kripke) −0.99 4.04 5.47 0.40
ActiGraph Link (Sadeh) −4.03 5.25 7.58 0.35
Actiwatch Spectrum 5.86 5.86 7.10 0.33
Oura Ring 2.89 3.93 4.86 0.61
Sleep onset latency (min) ActiGraph Link (Cole–Kripke) −4.16 5.00 6.89 0.45
ActiGraph Link (Sadeh) −1.53 3.92 5.44 0.59
Actiwatch Spectrum −7.87 7.87 10.52 0.13
Oura Ring −1.46 5.31 8.86 0.30
WASO (min) ActiGraph Link (Cole–Kripke) 9.61 20.61 29.25 0.47
ActiGraph Link (Sadeh) 23.37 27.50 41.00 0.43
Actiwatch Spectrum −23.65 24.13 30.42 0.44
Oura Ring −14.10 20.75 23.67 0.67
Total sleep time (min) ActiGraph Link (Cole–Kripke) −5.45 21.84 29.68 0.39
ActiGraph Link (Sadeh) −21.84 28.39 41.10 0.34
Actiwatch Spectrum 31.52 31.51 38.19 0.39
Oura Ring 15.56 21.21 26.22 0.62

Displayed are the mean error, mean absolute error (MAE), root mean squared error (RMSE), and R2.