Table 3.

Normalised air kerma and transmission measurements of the primary X-ray beam before and after transmission through an ICU cubicle glass door

Kilovoltage (kV) Air Kerma (uGy/mAs) Percentage transmission (%)
Before glass After glass
90 193 80 41
100 251 111 44
110 302 140 46
120 358 175 49

No inverse square law correction applied