Table 5.

Normalised scatter air kerma measurements at different distances from an anthropomorphic phantom for the standard technique for an erect and supine chest X-ray using a maximum 120 kV

Measurement direction Scatter air kerma (µGy/mAs)
Erect Supine
1 m 2 m 3 m 1 m 2 m 3 m
Side scatter 0.16 0.03 0.02 0.23 0.07 0.03
45 degrees 0.14 0.04 0.02 0.31 0.08 0.03
Back scatter 0.06 0.03 NA 0.27 0.06 NA
Forward scatter 0.25 0.06 NA 0.27 0.07 0.03

Due to the room set-up measurements in some directions could not be made at 3 m