Published in final edited form as: IEEE Trans Biomed Eng. 2020 Jan 16;67(9):2628–2637. doi: 10.1109/TBME.2020.2967033

TABLE I.

Mean thickness, incident angles and the percentage of phase difference less than 45° and 20° of all the measured spots on skulls 1-4 (ntotal = 784).

Incident Angle (°) Percentage (%) of Phase Shift Difference

#Skull Thickness (mm) Outer Surface Inner Surface <45° <20°

Resonance vs Hydrophone CT-based vs Hydrophone Resonance vs Hydrophone CT-based vs Hydrophone
Sk1 6.7 ± 0.6 5.5 ± 3.2 7.8 ± 4.5 67.4 59.1 36.2 22.8
Sk2 7.0 ± 1.5 4.5 ±2.5 9.0 ± 5.6 86.1 76.2 45.0 31.1
Sk3 5.3 ± 0.6 5.0 ± 2.8 9.1 ± 5.4 82.4 91.2 42.0 65.5
Sk4 8.5 ± 0.9 5.5 ± 3.2 9.2 ± 4.4 67.9 81.8 34.5 32.5

All 4 skulls 74.0 74.5 38.5 35.1

All 4 Skulls (incident angle < 5°) 80.9 76.2 45.4 37.2