TABLE 2│.
Troubleshooting table
| Step | Problem | Possible reason | Solution |
|---|---|---|---|
| 68 | Trapped air bubbles | Sudden injection of IP buffer into the hydrophobic PDMS device | Set the flow rate at 50 μl/min. Close the sieve valve. Wait for 5 s, and then open the valve to flush the air bubble out with the help of pressure generated in the chamber. Alternatively, press the chamber with a pair of tweezers to push the bubble out. |
| 78 | Air bubbles in the chamber after chromatin solution loading | The air bubble that separates IP buffer and chromatin solution is too large. | C-Flex tubing in long-tubing assembly needs to be changed periodically (roughly after 20 experiments) because it may expand over time and hold large air bubble during solution loading. |
| 100 | No visible pellet at the bottom of the microtube | Low concentration of ammonium acetate or ethanol | Be sure to prepare 5 M ammonium acetate solution in ultrapure water accurately. Use high quality glycogen and ethanol. |
| 107 | Low enrichment (<10 when primers in Table 1 are used) for ChIP DNA detected by qPCR | Chromatin fragmentation is not optimal | Adjust conditions for cross-linking and sonication to achieve 200–500 bp fragment size. Keep the temperature of the sample around 4 °C during sonication. For MNase digestion-based fragmentation, try different concentrations of MNase to achieve 170–540 bp fragment size. |
| The quality of antibody is not good. | Test antibodies from other batches or suppliers. | ||
| Problem with the reagents used in the experiment. | Make sure to add protease inhibitors as instructed in the protocol. | ||
| Oscillatory washing conditions are not optimal. | Adjust the time and pressure associated with washing. Generally, increased washing improves enrichment but decreases DNA amount. Ideal percent input is lower than 1% for negative loci, and 10%-40% for positive loci. | ||
| Primer design is not optimal | Design alternative primer sets. | ||
| Selected loci are not optimal | Target other loci in the genome. Order commercial primer sets if they are available. | ||
| 137 | More than 16 cycles are needed in library amplification when the amount of ChIP DNA is > 10 pg | Too much DNA is lost during SPRI beads cleanup steps. | Perform the SPRI cleanup with extra attention and use correct buffer ratio. Be sure to dry the beads after 80% (vol/vol) ethanol washing and also avoid over-drying the beads. |
| 143 | A large amount of adapter dimer in the library (~120 bp) | Insufficient size-selection step | Do the size-selection cleanup with accurate SPRI-to-buffer ratio. |