. 2020 Jul 9;25(27):2001223. doi: 10.2807/1560-7917.ES.2020.25.27.2001223

Table 2. Factors influencing the performance of Coronavirus disease (COVID-19) outbreak preparedness EQA participants, logistic regressions on combined responses reported for all samples in the panel, April/May 2020 (n = 4,168).

Possible technical influence factors (grouped by methods) Total responses per group Percentage correct results p value Odds ratio 95% confidence interval
% Responses (n)
Extraction method (overall) 4,168 96.3 4,012  < 0.001*
Any other method (incl. seven in-house) 1,152 94.6 1,090 Reference level
Abbott m2000sp 168 99.4 167 0.026* 9.50 1.31–68.97
BD MAX ExK TNA-3 144 90.3 130 0.040* 0.53 0.29–0.97
Roche Cobas Omni 304 99.3 302 0.003* 8.59 2.09–35.32
PSS MagDEA Dx 72 100 72 0.974 NA NA
Roche MagNA Pure 96 DNA and Viral 336 96.7 325 0.119 1.68 0.88–3.23
Promega Maxwell RSC (Viral/Blood) 176 95.5 168 0.644 1.19 0.56–2.54
BioMérieux NucliSENS easyMAG 544 98.4 535 < 0.001* 3.38 1.67–6.86
Qiagen QIAamp Viral RNA Mini Kit 184 98.9 182 0.023* 5.18 1.26–21.35
Qiagen QIAsymphony DSP Virus 144 97.2 140 0.189 1.99 0.71–5.56
Seegene STARMag Universal Cart. 360 98.1 353 0.009* 2.87 1.30–6.32
Cepheid Xpert Xpress SARS-CoV-2 376 94.4 355 0.880 0.96 0.58–1.60
Not specified 208 92.8 193 0.295 0.73 0.41–1.31
Amplification method (overall) 4,168 96.3 4,012  < 0.001*
Any real-time in-house PCR 952 96.7 921 Reference level
Abbott RealTime SARS-CoV-2 136 99.3 135 0.138 4.54 0.62–33.53
Altona RealStar SARS-CoV-2 RT-PCR 224 99.1 222 0.072 3.74 0.89–15.73
Cepheid Xpert Xpress SARS-CoV-2 408 94.9 387 0.098 0.62 0.35–1.09
Certest VIASURE SARS-CoV-2 136 91.9 125 0.008* 0.38 0.19–0.78
Certest VIASURE SARS-CoV-2 S gene 152 90.8 138 < 0.001* 0.33 0.17–0.64
Elitech GeneFinder COVID-19 Plus 200 95.0 190 0.230 0.64 0.31–1.33
Roche cobas SARSCoV2 296 99.3 294 0.029* 4.94 1.18–20.77
Seegene Allplex 2019-nCoV 552 98.9 546 0.013* 3.06 1.27–7.39
TIB MOLBIOL LightMix Modular 216 98.6 213 0.153 2.39 0.72–7.89
Any other commercial test kit 896 93.9 841 0.004* 0.52 0.33–0.81

NA: not applicable.

* Statistically significant; p ≤ 0.05 (for details see text).